Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Diagnostic International, Ltd.: AxSYM Digoxin II Reagent Pack, Disk Version 1. (3D53-01 or 3D53-02) The AxSYM Digoxin II assay is a Microparticle Enzyme Immunoassay (MEIA) for the quantitative measurement of digoxin, a cardiovascular drug, in serum or plasma. The measurements obtained are used in the treatment of digoxin overdose and monitoring levels of digoxin to ensure appropriate therapy.

Recall number
Z-0297-2009
Recalling firm
Abbott Diagnostic International, Ltd.
Product
AxSYM Digoxin II Reagent Pack, Disk Version 1. (3D53-01 or 3D53-02) The AxSYM Digoxin II assay is a Microparticle Enzyme Immunoassay (MEIA) for the quantitative measurement of digoxin, a cardiovascular drug, in serum or plasma. The measurements obtained are used in the treatment of digoxin overdose and monitoring levels of digoxin to ensure appropriate therapy.
Status
Terminated
Initiated
2008-05-22
Posted
2008-11-21
Terminated
2011-09-09
Reason
AxSYM TDM/ Transplant Assay Disks 3D53-03 and higher do not contain the AxSYM Digoxin II Assay file. The instrument might require installation of the AxSYM Digoxin II assay file.
Root cause
Software change control
Action
A Product Information letter, dated May 22, 2008, was issued to customers. It requested that one of the following two actions be taken if the AxSYM Digoxin II assay file is not installed on your AxSYM instrument and you want to run the AxSYM Digoxin II assay. If you have one of the previously manufactured software disks listed, [AxSYM Digoxin II Assay Disk, LN 2C43-01, or higher; or AxSYM TDM/Transplant Assay Disk Version 1.0 (3D53-01) or Version 2.0 (3D53-02)], install the AxSYM Digoxin II assay file. Do not install any other assay file. If you do not have one of the assay disks listed, contact your local Customer Service representative to have this assay file installed or install the file from a previous version if you have it on hand.
Quantity
9069 kits
Distribution
Worldwide Distribution --- USA including states of AL, AZ, AR, CA, CT, FL, GA, IL, KS, MD, MA, MI, MN, MS, MO, NV, NJ, NY, ND, OK, OR, PA, RI, TX, VA, WA, Hawaii, and Puerto Rico, and countries of Canada, Mexico, Guatemala, Costa Rica, Venezuela, Colombia, Ecuador, Peru, Chile, Uruguay, Argentina, Puerto Rico, Germany, Japan, Singapore, Hong Kong, Taiwan, Thailand, South Korea, Australia, New Zealand, Curacao, and Panama.
Lot, serial or model codes
Lot 61275Q100, exp. date: 09/09/08; Lot 58167Q100, exp. date: 08/30/08; Lot 56039Q101, exp. date: 07/13/08; and Lot 56039Q100, exp. date: 07/13/08.
510(k) numbers
["K953718"]
PMA numbers
not on file
Product code
KXT
Firm FEI
2623532
Firm city
Barceloneta
Firm state
PR
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register