Siemens Medical Solutions USA, Inc.: Siemens, ACUSON Antares, Ultrasound System (Standard and Premium Edition), 5.0 Systems with software version 200.0.054, P/N : 10032747 and 10032746 standard, 10032746, 10037591, and 10038202 for premium edition, Siemens, Ultrasound Division, Mountain View, CA 94039

Recall number
Z-0297-2008
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens, ACUSON Antares, Ultrasound System (Standard and Premium Edition), 5.0 Systems with software version 200.0.054, P/N : 10032747 and 10032746 standard, 10032746, 10037591, and 10038202 for premium edition, Siemens, Ultrasound Division, Mountain View, CA 94039
Status
Terminated
Initiated
2007-07-10
Posted
2008-01-09
Terminated
2010-12-20
Reason
Image errors: A software problem results in on-screen indications (graphics) that lead the user to believe that the patient's right and left and the transducer orientation are oriented the same. This is not the case and the actual patient's right and left are reversed from the on-screen indicators.
Root cause
Software design
Action
The firm sent a recall letter, Urgent Medical Device Notification, to consignees on 07/10/2007, to be followed by a service visit by Siemens personnel for a mandatory software update to be installed. Short term the users were instructed to avoid increasing the SieClear compounding levels to an Advanced SieClear compounding setting after an L/R invert action with Biopsy guidelines on-screen.
Quantity
418 units (Z-0297-0298-2008)
Distribution
Worldwide: Product was distributed to medical facilities Nationwide including the states of OH, FL, IL, NY, CA, NH, TX, NJ, PA, MD, MO, GA, WA, WI, NE, WV, HI, MA, SC, MI, VA, AZ, UT, NH, RI, KY, SD, and IN, and internationally to Canada, Germany, Australia, Sweden, Saudi Arabia, Chile, South Korea, Mexico, France, UK, Taiwan, Singapore, Denmark, Brazil, New Zealand, Ireland, Belgium, Norway, Serbia, Vietnam, UK, Poland, China, Spain, South Africa, Austria, Croatia, Russia, Greece, Trinidad and Tobago, United Arab Emirates, India, Japan, Malaysia, Turkey, Thailand, Norway, Slovakia, Romania, Netherlands, Switzerland, Portugal, Italy, and Venezuela.
Lot, serial or model codes
Serial numbers : 114792, 114819, 114822, 114827, 114829, 114839, 114843, 114844, 114845, 114846, 114847, 114848, 114849, 114851, 114859, 114860, 114861, 114863, 114864, 114871, 117874, 114876, 114880, 114881, 114882, 114885, 114890, 114891, 114894, 114895, 114909, 114910, 114912, 114918, 114921, 114927, 114928, 114930, 114931, 114935, 114941, 114944, 114947, 114948, 114949, 114953, 114957, 114958, 114959, 114960, 114962, 114963, 114971, 114977, 114978, 114730, 114735, 114745, 114761, 114781, 114785, 114826, 114865, 114868, 114869, 114896, 114936, 5000, 5001, 112919, 112925, 112971, 112980, 112981, 112984, 112989, 112992, 113005, 113009, 113013, 113015, 113037, 113040, 113041, 113042, 113045, 113048, 113051, 113052, 113053, 113055, 113058, 113061, 113067, 113068, 113075, 113076, 113079, 113080, 113101, 113102, 113105, 113143, 113145, 113147, 113204, 113225, 113257, 113259, 113302, 113407, 113411, 113414, 113462, 113469, 113566, 113682, 113706, 113715, 113759, 113767, 113803, 113825, 113826, 113827, 113828, 113829, 113831, 113832, 113834, 113835, 113841, 113843, 113846, 113847, 113848, 113849, 113850, 113851, 113852, 113856, 113863, 113873, 113900, 113917, 113921, 113945, 113947, 113950, 113957, 113980, 113986, 114014, 114043, 114052, 114061, 114164, 114172, 114176, 114186, 114233, 114254, 114255, 114271, 114274, 114313, 114351, 114356, 114363, 114406, 114461, 114474, 114475, 114481, 114482, 114493, 114536, 114573, 114583, 114584, 114590, 114596, 114587, 114602, 114612, 114614, 114616, 114617, 114621, 114640, 114644, 114647, 114652, 114655, 114661, 114665, 114669, 114671, 114675, 114687, 114690, 114694, 114699, 114703, 114706, 114716, 114721, 114728, 114734, 114732, 114740, 114742, 114743, 114744, 114749, 114751, 114753, 114754, 114755, 114757, 114759, 114760, 114762, 114764, 114768, 114770, 114771, 114773, 114775, 114777, 114782, 114789, 114790, 114791, 114792, 114793, 114794, 114796, 114799, 114802, 114805, 114807, 114808, 114809, 114810, 114811, 114813, 114818, 114819, 114820, 114821, 114822, 114825, 114827, 114828, 114829, 114831, 114833, 114834, 114835, 114836, 114838, 114842, 114843, 114844, 114844, 114845, 114846, 114847, 114848, 114849, 114850, 114851, 114853, 114857, 114859, 114860, 114861, 114863, 114864, 114866, 114870, 114873, 114878, 114886, 114899, 114901, 114908, 114914, 114916, 114920, 114926, 114937, 114939, 114940, 114943, 114945, 114950, 114955, 114973, 113322, 113904, 114228, 114415, 114524, 114528, 114566, 114568, 114623, 114630, 114632, 114635, 114636, 114645, 114646, 114651, 114672, 114685, 114717, 114723, 114724, 114725, 114726, 114727, 114729, 114733, 114736, 114737, 114739, 114741, 114746, 114747, 114748, 114750, 114752, 114758, 114769, 114765, 114766, 114767, 114779, 114780, 114783, 114784, 114786, 114788, 114795, 114797, 114798, 114800, 114801, 114803, 114804, 114806, 114812, 114814, 114816, 114823, 114824, 114830, 114832, 114837, 114840, 114841, 114852, 114854, 114855, 114858, 114862, 114867, 114868, 114869, 114872, 114875, 114877, 114879, 114887, 114888, 114889, 114892, 114893, 114897, 114898, 114900, 114902, 114903, 114904, 114905, 114907, 114911, 114913, 114915, 114919, 114922, 114923, 114924, 114925, 114929, 114933, 113249001, 113344001, 113458001, 113593001, 113731001, 113887001, and 113159001
510(k) numbers
["K063803"]
PMA numbers
not on file
Product code
IYN
Firm FEI
2936884
Firm city
Mountain View
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register