Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Laboratories HPD/ADD/GPRD: Abbott Commander Flexible Pipetting Center (FPC) Operations Manual, list 6A97-27; Abbott Laboratories, Abbott Park, IL 60064

Recall number
Z-0297-04
Recalling firm
Abbott Laboratories HPD/ADD/GPRD
Product
Abbott Commander Flexible Pipetting Center (FPC) Operations Manual, list 6A97-27; Abbott Laboratories, Abbott Park, IL 60064
Status
Terminated
Initiated
2003-12-02
Posted
2004-07-20
Terminated
2004-07-20
Reason
The Operations Manual has the wrong Farenheit temperature range listed (102-122) instead of (104-122) for the preparation of the Terg-A-Zyme solution used to clean pipettes of the Abbott Commander Flexible Pipetting Center.
Root cause
Other
Action
Abbott sent 'Important Product Information' letters dated 12/2/03 to the end user accounts, advising them of the error in conversion of the Celsius temperature range to Farenheit in the first step in preparing the Terg-A-Zyme cleaning solution. The letter listed the entire corrected solution preparation procedure and advised the accounts to retain the letter with the Abbott FPC Operations Manual for future reference.
Quantity
not on file
Distribution
Nationwide and internationally through Abbott affiliates in Colombia, Brazil, Puerto Rico, Japan, Singapore, Hong Kong and Germany.
Lot, serial or model codes
all revision numbers, all versions
510(k) numbers
["K894069"]
PMA numbers
not on file
Product code
JQW--
Firm FEI
1415939
Firm city
Abbott Park
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register