FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: ATS 3000 Automatic Tourniquet System w/HOSES AND LOP SENSOR Intended to be used by qualified medical professionals to temporarily occlude blood flow in a patient's extremities during surgical procedures on those extremities.

Recall number
Z-0296-2013
Recalling firm
Zimmer, Inc.
Product
ATS 3000 Automatic Tourniquet System w/HOSES AND LOP SENSOR Intended to be used by qualified medical professionals to temporarily occlude blood flow in a patient's extremities during surgical procedures on those extremities.
Status
Terminated
Initiated
2012-09-05
Posted
2012-11-15
Terminated
2014-04-17
Reason
Zimmer is conducting a voluntary removal of some Zimmer A.T.S. 3000 tourniquets after receiving complaints of both out-of-box failures and failures during surgical procedures on some of the affected units. During failure the unit will alarm, lock into a non-operational mode, and the display screen will go blank. Based on investigation, this problem appears to be due to a counterfeit chip on the
Root cause
Counterfeit
Action
Zimmer sent an Urgent Device Removal Notice letter dated September 5, 2012, to all affected customers. Zimmer Distributors and Sales Leadership were notified via E-mail of the pending recallthe product under recall beginnig September 11, 2012. The notification listed the reason for the recall, risks involved, and requested return of the unit for repair To: Zimmer Surgical, Attn: QAlRA Dept. - Recall, 200 West Ohio Avenue, Dover, Ohio 44622 USA The notice asked to ensure all users be made aware of the recall, remove the affected ATS 3000 tourniquest sytem for repair and return the fax back form/acknowledgement certification to 1-888-429-7380 or email it to zimmer3611@stericycle.com. Questions should be directed to 1-8889435167 between 8 a.m. and 5 p.m. EDT. For questions regarding this recall call 574-372-4807.
Quantity
204
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Belgium, Czech Republic, Egypt, France, Germany, Morocco, Netherlands, Romania, Saudi Arabia, Spain, Switzerland, Turkey, and UAE.
Lot, serial or model codes
Catalog 60-3000-101-00,  Serial # 3005IACJ, 3006JAEP, 3007CADR, 3007EACT, 3007EAKT, 3007GAGR, 3007HABK, 3007KAAR, 3007KAET, 3008CADE, 3008HAHC, 3008LACK, 3009AACN, 3009HABA, 3009LAEF, 3010CABH, 3010DAHG, 3010GAAC, 3010GAAJ, 3011LABL, 3011LABN, 3011LABQ, 3011LACC, 3011LACE, 3011LACF, 3011LACG, 3011LACH, 3011LACJ, 301 1LACK, 3011LACL, 3011LACM, 3011LACN, 3011LACP, 3011LACQ, 3011LACR, 3011LACS, 3011LACT, 3011LACU, 3011LACW, 3011LADA, 3011LADE, 3011LADF, 3011LADG, 3011LADH, 3011LADJ, 3011LADK, 3011LADM, 3011LADP, 3011LADQ, 3011LADR, 3011LADS, 3011LADT, 3011LADU, 3011LADW, 3011LAEA, 3011LAEB, 3011LAEC, 3011LAED, 3011LAEE, 3011LAEF, 3011LAEH, 3012AAAG, 3012AAAK, 3012AAAM, 3012AAAN, 3012AAAP, 3012AAAQ, 3012AAAR, 3012AAAS, 3012AAAT, 3012AAAU, 3012AAAW, 3012AABA, 3012AABB, 3012AABC, 3012AABE, 3012AABG, 3012AABH, 3012AABK, 3012AABL, 3012AABM, 3012AABN, 3012AABQ, 3012AABR, 3012AABS, 3012AABU, 3012AACA, 3012AACB, 3012AACC, 3012AACD, 3012AACE, 3012AACH, 3012AACJ, 3012AACM, 3012AACN, 3012AACP, 3012AACQ, 3012AACR, 3012AACS, 3012AACT, 3012AACW, 3012AADA, 3012AADB, 3012AADC, 3012AADD, 3012AADE, 3012AADF, 3012AADG, 3012BAAA, 3012BAAB, 3012BAAC, 3012BAAD, 3012BAAE, 3012BAAF, 3012BAAG, 3012BAAH, 3012BAAJ, 3012BAAK, 3012BAAL, 3012BAAM, 3012BAAN, 3012BAAP, 3012BAAQ, 3012BAAR, 3012BAAS, 3012BAAT, 3012BAAU, 3012BAAW, 3012BABA, 3012BABB, 3012BABC, 3012BABD, 3012BABE, 3012BABF, 3012BABG, 3012BABH, 3012BABJ, 3012BABK, 3012BABL, 3012BABM, 3012BABN, 3012BABP, 3012BABQ, 3012BABR, 3012BABS, 3012BABT, 3012BABU, 3012BABW, 3012BACE, 3012BACG, 3012BACH, 3012BACJ, 3012BACK, 3012BACL, 3012BACM, 3012BACN, 3012BACP, 3012BACQ, 3012BACR, 3012BACS, 3012BACT, 3012BACU, 3012BACW, 3012BADA, 3012BADB, 3012BADC, 3012BADD, 3012BADG, 3012BADH, 3012BADJ, 3012BADK, 3012BADL, 3012BADN, 3012BADP, 3012BADR, 3012BADS, 3012BADU, 3012BADW, 3012BAEA, 3012BAEB, 3012BAEC, 3012BAEH, 3012BAEJ, 3012BAEL, 3012BAEM, 3012BAEN, 3012BAEQ, 3012BAER  French Catalog 60-3000 301-02 Serial # 3012 AAAE German Catalog 60-300060-301-03, Serial # 3012AAAF Spanish Catalog 60-3000-301-11, Serial # 3012AAAB, 3012AAAC, 3012AACG, 3012BACA, 3012BACB, 3012BACC, 3012BACD Dutch 60-3000-301-12, Serial # 3012AAAA, 3012AAAD, 3012EACA
510(k) numbers
not on file
PMA numbers
not on file
Product code
KCY
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register