Skip to the content

FDA Medical Device Recalls (openFDA)

Polymedco, Inc: POLYMEDCO ULTRA-CRP LATEX ENHANCED ASSAY, Catalog No. CPU400. Contents: 1 x 50 ML R1, 2 x 6 ML R2. R1 is the HIGH SENSITIVE CRP Buffer IVD (50 ml). R2 is the HIGH SENSITIVE CRP Latex IVD (6 ml). For In Vitro Diagnostic (IVD) use only. COMMON/USUAL NAME: High-Sensitive C-Reactive Protein --- CLASSIFICATION NAME: C-reactive Protein, Antigen, Antiserum

Recall number
Z-0296-04
Recalling firm
Polymedco, Inc
Product
POLYMEDCO ULTRA-CRP LATEX ENHANCED ASSAY, Catalog No. CPU400. Contents: 1 x 50 ML R1, 2 x 6 ML R2. R1 is the HIGH SENSITIVE CRP Buffer IVD (50 ml). R2 is the HIGH SENSITIVE CRP Latex IVD (6 ml). For In Vitro Diagnostic (IVD) use only. COMMON/USUAL NAME: High-Sensitive C-Reactive Protein --- CLASSIFICATION NAME: C-reactive Protein, Antigen, Antiserum
Status
Terminated
Initiated
2003-12-03
Posted
2004-01-06
Terminated
2005-04-20
Reason
Firm was notified by the foreign manufacturer that certain lots of Ultra-CRP are prone to prozone (a condition of antibody antigen excess) and do not meet the labeled specifications.
Root cause
Other
Action
By phone on 12/3/03 and by mail on 12/5/03.
Quantity
584 Kits
Distribution
Medical facilities, physicians, diagnostic laboratories, and distributors located nationwide.
Lot, serial or model codes
Lot numbers 1155J (EXP. 09/04) and 1388J (EXP. 10/04)
510(k) numbers
["K001009"]
PMA numbers
not on file
Product code
DCK
Firm FEI
1000119795
Firm city
Cortlandt Manor
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register