FDA Medical Device Recalls (openFDA)
Ethicon Endo-Surgery Inc: Endoscopic Articulating Linear Cutter, ATB35. Standard 35 MM. The ENDOPATH ETS-Flex Endoscopic Articulating Linear Cutter has application in general, gynecologic, urologic, and thoracic surgery for transection, resection, and/or creation of anastomoses
- Recall number
- Z-0295-2017
- Recalling firm
- Ethicon Endo-Surgery Inc
- Product
- Endoscopic Articulating Linear Cutter, ATB35. Standard 35 MM. The ENDOPATH ETS-Flex Endoscopic Articulating Linear Cutter has application in general, gynecologic, urologic, and thoracic surgery for transection, resection, and/or creation of anastomoses
- Status
- Terminated
- Initiated
- 2016-07-15
- Posted
- 2016-10-20
- Terminated
- 2017-02-28
- Reason
- Quality control identified a component issue where the pinion gear in the device could fail under extreme use cases. If this condition occurs, staples will be formed past the cut line and the device can be opened and removed from the patient. The firing stroke may be interrupted and the knife may not fully return to the home position potentially exposing the healthcare professional to a sharps injury.
- Root cause
- Device Design
- Action
- Ethicon sent an Urgent Medical Device Recall letter dated July 15, 2016, to all affected consignees via UPS next day mail. The letter identified the product, the problem, and the action to be taken by the customer. Consignees were instructed to examine their inventory immediately and quarantine any affected product and return to ENDOPATH. If product subject to the recall was further distributed consignees were instructed to notify the other facility and arrange for return of the affected product. Customer were instructed to complete the Business Reply Form confirming receipt of this letter and fax to Stericycle at 1-888-228-5719 or email to ethicon5182@stericycle.com. Customers with questions were instructed to contact the Ethicon Customer Support Center at 1-877-384-4266. For questions regarding this recall call 513-337-3582.
- Quantity
- 105
- Distribution
- Worldwide Distribution - US including AL, AR, AZ, CA, CO,CT, DC, FL,GA, IA,IL,KS,KY,LA,MA,MD, ME, MI,MN,MO, MS, MT,NC, ND, NH,NJ, NM, NV, NY, OH, RI, TN, TX, VA, VT, WA, WI and Internationally to Brazil and Canada.
- Lot, serial or model codes
- N91L48, N91L49, N91R40, N91R41, N91U0W, N91U0X
- 510(k) numbers
- ["K111111"]
- PMA numbers
- not on file
- Product code
- GDW
- Firm FEI
- 1527736
- Firm city
- Blue Ash
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28