Skip to the content

FDA Medical Device Recalls (openFDA)

Argon Medical Devices, Inc: (1) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Small Straight, REF 046-1-200, 2 clamps/package, 5 packages/box, 10 boxes/case, Sterile, Rx; (2) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Medium Straight, REF 046-1-300, 2 clamps per package, 5 packages/ shelf box, 10 shelf boxes/case (for a total of 50/2-clamp packages/case), Sterile, Rx; (3) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Medium Angled, REF 046-1-301, 2 clamps per package, 6 packages/box, 50 boxes/case, Sterile, Rx; and (4) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Large Angled, REF 046-1-401, 2 clamps per package, 5 packages/box, 50 boxes/case. The firm name on the label is Argon Medical Devices, Inc., Athens, TX.

Recall number
Z-0293-2020
Recalling firm
Argon Medical Devices, Inc
Product
(1) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Small Straight, REF 046-1-200, 2 clamps/package, 5 packages/box, 10 boxes/case, Sterile, Rx; (2) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Medium Straight, REF 046-1-300, 2 clamps per package, 5 packages/ shelf box, 10 shelf boxes/case (for a total of 50/2-clamp packages/case), Sterile, Rx; (3) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Medium Angled, REF 046-1-301, 2 clamps per package, 6 packages/box, 50 boxes/case, Sterile, Rx; and (4) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Large Angled, REF 046-1-401, 2 clamps per package, 5 packages/box, 50 boxes/case. The firm name on the label is Argon Medical Devices, Inc., Athens, TX.
Status
Terminated
Initiated
2018-10-16
Posted
not on file
Terminated
2021-06-15
Reason
The seal formed on the edge of the pouch may be incomplete, which could result in a loss of sterility to the product.
Root cause
Equipment maintenance
Action
The recalling firm began issuing letters dated 10/16/2018 on 10/16/2018 via registered mail identifying the issue and the action to be taken by the customer.
Quantity
Total number in recall for all products in dist: 49,130 units
Distribution
Distribution was nationwide. There was military/government distribution. Foreign distribution was made to Canada, Australia, the Netherlands, and Spain.
Lot, serial or model codes
(1) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Small Straight, REF 046-1-200 - Lot number 11230432;  (2) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Medium Straight, REF 046-1-300 - Lot numbers 11226690 and 11231581  (3) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Medium Angled, REF 046-1-301 - Lot number 11227669;  (4) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Large Angled, REF 046-1-401 - Lot numbers 11228371 and 11230677.
510(k) numbers
["K771701"]
PMA numbers
not on file
Product code
DRS
Firm FEI
1625425
Firm city
Athens
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register