Skip to the content

FDA Medical Device Recalls (openFDA)

Toshiba American Med Sys Inc: Toshiba-Interventional Angiography System, Infinix i, Model DFP-8000D, equiped with Digital Radiography System with Software version 3.4er000, Toshiba America Medical Systems, Inc., Tustine, CA 92780

Recall number
Z-0293-2008
Recalling firm
Toshiba American Med Sys Inc
Product
Toshiba-Interventional Angiography System, Infinix i, Model DFP-8000D, equiped with Digital Radiography System with Software version 3.4er000, Toshiba America Medical Systems, Inc., Tustine, CA 92780
Status
Terminated
Initiated
2007-10-01
Posted
2007-11-27
Terminated
2012-04-05
Reason
System Lock-up: When performing DAS One Shot, the system may lockup due to a software timing error.
Root cause
Software design
Action
Notification letters (Urgent: Medical Device Correction-dated October 1, 2007) with an attached Customer Response form were sent to customers beginning on October 2, 2007. Customers were notified this action was being taken as a result of an investigation that resulted from a software issue that was found done by the manufacturer. Customers were informed that a Toshiba Customer Service Engineer would contact them to schedule a software upgrade when the software becomes available. If customers have any questions they were instructed to contact their or call (800) 421-1968 and ask for the Regulatory Affairs Department.
Quantity
23
Distribution
Nationwide-including states of AL, AR, FL, GA, MI, NY, OH, PA, TX & WV
Lot, serial or model codes
Serial numbers: B5522165, W1C0562187, W1C0572197, W1C0572202, W1C0622275, W1C0622270, A3632065, A3622066, W1C0456192, W1C0632284, W1D0722373, W1D0742386, B5512160, W1C0592213, W1C0592215, A4562099, W1C05Z2254, B4582113, W1C0642289, W1D0697015, W1C0552182, W1C05X2224, W1D0742387
510(k) numbers
["K052884"]
PMA numbers
not on file
Product code
MQB
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register