Skip to the content

FDA Medical Device Recalls (openFDA)

Ethicon Endo-Surgery Inc: REALIZE Adjustable Gastric Band Pak with Endoscopic Dissector and 15mm XCEL" trocar Product Code RLZB22DT, Lot #: G4RK31, G4RR27, G4T28C

Recall number
Z-0292-2011
Recalling firm
Ethicon Endo-Surgery Inc
Product
REALIZE Adjustable Gastric Band Pak with Endoscopic Dissector and 15mm XCEL" trocar Product Code RLZB22DT, Lot #: G4RK31, G4RR27, G4T28C
Status
Terminated
Initiated
2010-09-27
Posted
2010-11-02
Terminated
2011-11-18
Reason
The Tubing Strain Relief component of the Injection Port may detach from the Locking Connector, and migrate from its original position on the Locking Connector. This may result in: Tubing Strain Relief migrating on the Tubing; Kinking of the Tubing resulting in difficulty/inability to adjust the fluid volume within the Band; or a free foreign body in the abdominal wall or abdominal (peritoneal) ca
Root cause
Other
Action
Ethicon Endo-Surgery issued an Urgent : Devoced Recall Event 2240 letter dated October 8, 2010 to all consignees identifying the affected devices and actions to be taken. Consignees were instructed: DO NOTt use product from the affected lots, but implanted devices do not need to be removed as part of this recall. Examine inventory and remove the affected product. If they DO have the affected product, complete the Business Reply Form, keeping a copy for records, and place the original form in the box with the product to be returned, affix the pre-printed label, and return to Stericycle. If they DO NOT have the affected product, complete the Business Reply Form, and fax it to Stericycle. A redesigned replacement product will be sent out to consignees within 30-60 days of product return. Ethicon Endo-Surgery can be contacted at 513 337-8208 concerning this recall.
Quantity
not on file
Distribution
Worldwide Distribution: USA including the following states: AK, AL, AR, AZ, CA, CO, CT, Washington DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, TN, TX, UT, VA, VT, WA, WI and WV, and the following countries: ARAB EMIRATES, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CYPRUS, CZECH REPUBLIC, EGYPT, FRANCE, GERMANY, GREAT BRITAIN, GREECE, GUATEMALA, HONG KONG, HUNGARY, INDIA, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, LIBYA, MEXICO, NETHERLANDS, PAKISTAN, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, RUSSIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, TAIWAN, THAILAND, TURKEY and VENEZUELA.
Lot, serial or model codes
Product Code RLZB22DT, Lot #: G4RK31, G4RR27, G4T28C
510(k) numbers
not on file
PMA numbers
["P070009"]
Product code
LTI
Firm FEI
1527736
Firm city
Cincinnati
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register