Skip to the content

FDA Medical Device Recalls (openFDA)

W L Gore & Associates, Inc.: GORE TAG Conformable Thoracic Stent Graft with ACTIVE CONTROL System

Recall number
Z-0291-2021
Recalling firm
W L Gore & Associates, Inc.
Product
GORE TAG Conformable Thoracic Stent Graft with ACTIVE CONTROL System
Status
Open, Classified
Initiated
2020-09-09
Posted
not on file
Terminated
not on file
Reason
When the delivery catheter was attempted to be removed, there was an observed connection between the stent graft and the delivery catheter that prevented the delivery catheter from being withdrawn.
Root cause
Process control
Action
On 09/23/2020, the firm sent an "URGENT Medical Device Safety Correction" via First Class Postal (USPS) informing them that all lots of its Thoracic Stent Graft Systems may encounter difficulty withdrawing the delivery catheter due to a proximal stent apex being lodged with the delivery catheter leading olive. Customers are instructed to: Immediate Actions for the Recipient: - Take note of amendment/reinforcement of Instructions For Use (IFU) - Please respond to the enclosed acknowledgement - Please share this letter with others in your hospital or clinic as appropriate The Recalling Firm is providing physicians with this information so that appropriate risk-related decisions can be made with the patient when considering the Conformable Thoracic Stent Graft System. The Recalling Firm maintains its confidence in the safety and efficacy of the device and will not be removing the affected products from the market. There are no actions required for patients already implanted with the Thoracic Stent Graft. Please contact the Recalling Firm's Customer Service (email: MPDCustomerCare@wlgore.com or by phone at 800-528-8763) with any questions related to this letter.
Quantity
19013 units
Distribution
U.S. - Nationwide O.U.S. - Australia, Austria, Belgium, Croatia, Cypress, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Lebanon, Luxembourg, Monaco, Netherlands, Norway, New Zealand, Poland, Portugal, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom
Lot, serial or model codes
CATALOG NUMBERS TGM212110X* TGM212115X* TGM212120X* TGM262610X* TGM262615X* TGM262620X* TGM282810X* TGM282815X* TGM282820X* TGM313110X* TGM313115X* TGM313120X* TGM343410X* TGM343415X* TGM343420X* TGM373710X* TGM373715X* TGM373720X* TGM404010X* TGM404015X* TGM404020X* TGM454510X* TGM454515X* TGM454520X* TGM262110X* TGM312610X* TGMR313110X* TGMR313115X* TGMR313120X* TGMR312610X* TGMR373710X* TGMR373715X* TGMR373720X* TGMR404010X* TGMR404015X* TGMR404020X* *X is a placeholder for enumeration code: E=EMEA, Australia, New Zealand; J=Japan  All Lot Numbers
510(k) numbers
not on file
PMA numbers
["P040043"]
Product code
MIH
Firm FEI
2017233
Firm city
Flagstaff
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register