Skip to the content

FDA Medical Device Recalls (openFDA)

Orthofix Srl: Orthofix Power Drill Torque Limiter, Sterile, Rx Only, Single Use; Sold individually and in the Galaxy UNYCO Diaphyseal Tibia Sterile Kit, Galaxy UNYCO Mini Kit Instruments Sterile, Galaxy UNYCO Diaphyseal Femur Box, Galaxy UNYCO Knee Bridging Box, Galaxy UNYCO Ankle Bridging Box - Unilateral Frame, Galaxy UNYCO Ankle Bridging Box - Delta Frame, and Galaxy UNYCO Diaphyseal Tibia Box. Intended to be used for temporary bone stabilization in trauma and orthopedic procedures of the lower limb prior to definitive treatment.

Recall number
Z-0291-2018
Recalling firm
Orthofix Srl
Product
Orthofix Power Drill Torque Limiter, Sterile, Rx Only, Single Use; Sold individually and in the Galaxy UNYCO Diaphyseal Tibia Sterile Kit, Galaxy UNYCO Mini Kit Instruments Sterile, Galaxy UNYCO Diaphyseal Femur Box, Galaxy UNYCO Knee Bridging Box, Galaxy UNYCO Ankle Bridging Box - Unilateral Frame, Galaxy UNYCO Ankle Bridging Box - Delta Frame, and Galaxy UNYCO Diaphyseal Tibia Box. Intended to be used for temporary bone stabilization in trauma and orthopedic procedures of the lower limb prior to definitive treatment.
Status
Terminated
Initiated
2017-07-05
Posted
not on file
Terminated
2023-01-13
Reason
The power drill torque limiter may break during use in surgery application.
Root cause
Process change control
Action
Consignees were sent an Urgent Medical Device Recall Notification, dated June 28, 2017, informing them of the device correction. The letter instructed distributors to immediately identify affected product in warehouses, quarantine affected product, and communicate the amount on hand to Orthofix SRL for replacement. Any customers are to be forwarded the recall notice provided to the distributor. Distributors are asked to complete the Reply Form with the appropriate information and to return it to Orthofix by fax at +39-045-6719380 or by e-mail to customerservice@orthofix.it. Hospitals were instructed to identify affected product on-site, quarantine affected units, and communicate the quantity of affected product on-hand to their local distributor or to Orthofix. Affected product will be replaced by Orthofix.
Quantity
416 units
Distribution
Worldwide Distribution -- US, including the states of CA, FL, & PA; and, the countries of Australia, Germany, Finland, France, Italy, Lithuania, Netherlands, Singapore, South Africa, Sweden, Switzerland, & United Kingdom.
Lot, serial or model codes
Power Drill Torque Limiter: Product Code 99-93568; Lot (Exp Date) V1388296 (12/28/2019), V1388432 (12/29/2019), V1389857 (01/13/2020), V1400707 (05/18/2020), V1400940 (05/23/2020), V1410027 (09/01/2020); UDI: 18054242512318.   Galaxy UNYCO Diaphyseal Tibia Sterile Kit: Product Code 99-93506; Lot (Exp Date) V1389818 (01/12/2020), V1389819 (01/12/2020), V1389886 (02/13/2020), V1399881 (05/10/2020), V1400800 (05/22/2020), V1402023 (06/03/2020), V1405602 (07/17/2020), V1406422 (07/26/2020), V1410082 (09/02/2020), R1412578 (09/28/2020), V1414451 (10/19/2020), V1421785 (02/01/2021), V1425406 (03/13/2021), B1012462 (06/19/2021), B1012776 (06/26/2021), B1019431 (07/05/2021), B1031608 (08/05/2021), B1052165 (10/23/2021), B1049195 (10/25/2021), B1080028 (02/16/2022), B1099157 (04/08/2022), B1107271 (04/09/2022); UDI: 18054242512325.   Galaxy UNYCO Mini Kit Instruments Sterile: Product Code 99-93509; Lot (Exp Date) V1408711 (08/01/2020), V1410026 (09/01/2020), V1422854 (02/14/2021), V1427350 (04/09/2021), V1429217 (04/26/2021), B1083360 (02/20/2022); UDI: 18054242513629.   Galaxy UNYCO Diaphyseal Femur Box: Product Code 99-93790; Lot (Exp Date) B1019281 (04/09/2021); UDI: 1805424515159.   Galaxy UNYCO Knee Bridging Box: Product Code 99-93791; Lot (Exp Date) B1019282 (04/09/2021); UDI: 18054242515166.  Galaxy UNYCO Ankle Bridging Box - Unilateral Frame; Product Code 99-93792; Lot (Exp Date) V140597 (03/21/2020); UDI: 18054242513643.   Galaxy UNYCO Ankle Bridging Box - Delta Frame: Product Code 99-93793; Lot (Exp Date) V1409944 (03/09/2020); UDI: 18054242513650.   Galaxy UNYCO Diaphyseal Tibia Box: Product Code 99-93794; Lot (Exp Date) V1410194 (04/18/2020). UDI: 18054242513667.
510(k) numbers
["K142052","K153233"]
PMA numbers
not on file
Product code
KTT
Firm FEI
3002807827
Firm city
Bussolengo
Firm state
not on file
Firm country
Italy

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register