FDA Medical Device Recalls (openFDA)
W L Gore & Associates, Inc.: GORE TAG CONFORMABLE Thoracic Stent Graft with ACTIVE CONTROL System
- Recall number
- Z-0290-2021
- Recalling firm
- W L Gore & Associates, Inc.
- Product
- GORE TAG CONFORMABLE Thoracic Stent Graft with ACTIVE CONTROL System
- Status
- Open, Classified
- Initiated
- 2020-09-09
- Posted
- not on file
- Terminated
- not on file
- Reason
- Due to secondary deployment fiber being attached to secondary deployment handle, completion of deployment via the deployment line access hatch was not possible
- Root cause
- Unknown/Undetermined by firm
- Action
- On 09/23/2020, the firm sent an "URGENT Medical Device Safety Correction" via First Class Postal (USPS) informing them that all lots of its Thoracic Stent Graft Systems may encounter inability to complete secondary deployment and that the Recalling Firm will be updating it Instruction for Use to include a new Warning: Secondary deployment events have occurred where the device does not open to full diameter after Secondary Deployment Handle actuation, and the deployment line is not accessible via the Deployment Line Access Hatch. If this event occurs, Gore recommends using the GORE Tri-Lobe Balloon Catheter to expand the device from trailing end to leading end after complete delivery system component removal. (Please refer to the GORE Tri-Lobe Balloon Catheter Instructions for Use for pertinent recommended volume, directions and warnings.) Ballooning of a non-fully expanded stent graft may lead to improper placement of the stent graft and/or branch vessel occlusion or obstruction. Use of an occlusion balloon may lead to device distal displacement during deployment (windsock effect) and has been observed to lead to branch vessel occlusion or obstruction. Customers are instructed to: -Take note of amendment/reinforcement of Instructions For Use (IFU) - Please respond to the enclosed acknowledgement - Please share this letter with others in your hospital or clinic as appropriate The Recalling Firm is providing physicians with this information so that appropriate risk-related decisions can be made with the patient when considering the Thoracic Stent Graft System. The Recalling Firm maintains its confidence in the safety and efficacy of the device and is not removing the affected products from the market. There are no actions required for patients already implanted with the Thoracic Stent Graft. Please contact the Recalling Firm's Customer Service (email: MPDCustomerCare@wlgore.com or by phone at 800-528-8763) with any questions related to this
- Quantity
- 13860 units
- Distribution
- U.S. - Nationwide O.U.S. - Australia, Austria, Belgium, Croatia, Cypress, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Lebanon, Luxembourg, Monaco, Netherlands, Norway, New Zealand, Poland, Portugal, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom
- Lot, serial or model codes
- CATALOG NUMBERS TGM212110X* TGM212115X* TGM212120X* TGM262610X* TGM262615X* TGM262620X* TGM282810X* TGM282815X* TGM282820X* TGM313110X* TGM313115X* TGM313120X* TGM343410X* TGM343415X* TGM343420X* TGM373710X* TGM373715X* TGM373720X* TGM404010X* TGM404015X* TGM404020X* TGM454510X* TGM454515X* TGM454520X* TGM262110X* TGM312610X* TGMR313110X* TGMR313115X* TGMR313120X* TGMR312610X* TGMR373710X* TGMR373715X* TGMR373720X* TGMR404010X* TGMR404015X* TGMR404020X* *X is a placeholder for enumeration code: E=EMEA, Australia, New Zealand; J=Japan All lot numbers
- 510(k) numbers
- not on file
- PMA numbers
- ["P040043"]
- Product code
- MIH
- Firm FEI
- 2017233
- Firm city
- Flagstaff
- Firm state
- AZ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28