Skip to the content

FDA Medical Device Recalls (openFDA)

Argon Medical Devices, Inc: (1) Argon Medical Devices Stopcocks, Clear 4-Way Stopcock w/Male Luer Lock, REF 041210002A, 25/box, Sterile, Rx; (2) Argon Medical Devices Stopcocks, 1-Way w/Male Luer Lock, REF 041216000A, 25/box, Sterile, Rx; (3) Argon Medical Devices Stopcocks, Double 4-Way Stopcock w/Male Luer Lock, REF 041220001A, 25/box, Sterile, Rx; (4) Argon Medical Devices Stopcocks, Clear 4-Way Stopcock w/Rotating Male Luer Lock, REF 041240002A, Sterile, Rx; (5) Argon Medical Devices Stopcocks, North Carolina Baptist Hospital, Winston-Salem, NC, REF 042597300A, 25/box, Sterile, Rx; and (6) Argon Medical Devices 3 Way Stopcock w/Rotating Adapter (1050psi), REF 196590, 25/box, Sterile, Rx. The firm name on the labels is Argon Medical Devices, Inc., Athens, TX.

Recall number
Z-0290-2020
Recalling firm
Argon Medical Devices, Inc
Product
(1) Argon Medical Devices Stopcocks, Clear 4-Way Stopcock w/Male Luer Lock, REF 041210002A, 25/box, Sterile, Rx; (2) Argon Medical Devices Stopcocks, 1-Way w/Male Luer Lock, REF 041216000A, 25/box, Sterile, Rx; (3) Argon Medical Devices Stopcocks, Double 4-Way Stopcock w/Male Luer Lock, REF 041220001A, 25/box, Sterile, Rx; (4) Argon Medical Devices Stopcocks, Clear 4-Way Stopcock w/Rotating Male Luer Lock, REF 041240002A, Sterile, Rx; (5) Argon Medical Devices Stopcocks, North Carolina Baptist Hospital, Winston-Salem, NC, REF 042597300A, 25/box, Sterile, Rx; and (6) Argon Medical Devices 3 Way Stopcock w/Rotating Adapter (1050psi), REF 196590, 25/box, Sterile, Rx. The firm name on the labels is Argon Medical Devices, Inc., Athens, TX.
Status
Terminated
Initiated
2018-10-16
Posted
not on file
Terminated
2021-06-15
Reason
The seal formed on the edge of the pouch may be incomplete, which could result in a loss of sterility to the product.
Root cause
Equipment maintenance
Action
The recalling firm began issuing letters dated 10/16/2018 on 10/16/2018 via registered mail identifying the issue and the action to be taken by the customer.
Quantity
Total number in recall for all products in dist: 49,130 units
Distribution
Distribution was nationwide. There was military/government distribution. Foreign distribution was made to Canada, Australia, the Netherlands, and Spain.
Lot, serial or model codes
(1) Argon Medical Devices Stopcocks, Clear 4-Way Stopcock w/Male Luer Lock, REF 041210002A-Lot numbers 11226656 and 11228887;  (2) Argon Medical Devices Stopcocks, 1-Way w/Male Luer Lock, REF 041216000A - Lot numbers 11226187, 11229619, and 11227967;  (3) Argon Medical Devices Stopcocks, Double 4-Way Stopcock w/Male Luer Lock, REF 041220001A - Lot numbers 11228888, 11230598;  (4) Argon Medical Devices Stopcocks, Clear 4-Way Stopcock w/Rotating Male Luer Lock, REF 041240002A - Lot number 11228245;  (5) Argon Medical Devices Stopcocks, North Carolina Baptist Hospital, Winston-Salem, NC, REF 042597300A - Lot number 11232206; and  (6) Argon Medical Devices 3 Way Stopcock w/Rotating Adapter (1050psi), REF 196590 - Lot numbers 11229800 and 11232485.
510(k) numbers
["K780271"]
PMA numbers
not on file
Product code
DRS
Firm FEI
1625425
Firm city
Athens
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06