Skip to the content

FDA Medical Device Recalls (openFDA)

Remel Inc: Remel Shigella sonnei Phase 1 & 2 Agglutinating Serum, Ref R30164201 ZH01. Product Usage: Suitable for the use in slide agglutination tests to identify Shigella cultures presumptively for epidemiological and diagnostic purposes.

Recall number
Z-0290-2018
Recalling firm
Remel Inc
Product
Remel Shigella sonnei Phase 1 & 2 Agglutinating Serum, Ref R30164201 ZH01. Product Usage: Suitable for the use in slide agglutination tests to identify Shigella cultures presumptively for epidemiological and diagnostic purposes.
Status
Terminated
Initiated
2017-12-08
Posted
not on file
Terminated
2020-07-22
Reason
The serum may fail to agglutinate within the specified minimum reaction time.
Root cause
Under Investigation by firm
Action
The recalling firm issued letters dated 12/8/2017 via regular mail. The letter identified the affected product, problem and actions to be taken. The risk to health is also explained. The customer is requested to notify any personnel who need to be aware of the potential for incorrect performance specification for the product. The customer is to inspect their stock and destroy any remaining inventory of the lots listed. They are to contact Customer Service or their local distributor for replacement. A response form was to be returned acknowledging receipt of the notice and disposal of the affected product. If the product was further distributed, the customer was requested to contact their downstream customers, advise them of the recall, and provide them a copy of the recall letter.
Quantity
237 units (single bottles)
Distribution
US Nationwide Distribution
Lot, serial or model codes
UDI 50560805009592. Kit lot #1465529, Bottle lot #1465528; Kit lot #2118879, Bottle lot #1847290; Kit lot #2173683, Bottle lot #1847290; Kit lot #2161363, Bottle lot #1847290; Kit lot #1984639, Bottle lot #1847290; Kit lot #1738564, Bottle lot#1738565; Kit lot #1971308, Bottle lot #1738565; Kit lot #1957305, Bottle lot #1738565; Kit lot #1947930, Bottle lot #1738565; and Kit lot #1847285, Bottle lot #1738565.
510(k) numbers
not on file
PMA numbers
not on file
Product code
GNB
Firm FEI
1924669
Firm city
Lenexa
Firm state
KS
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register