Medtronic Neurological: Medtronic Kinetra Dual Program Neurostimulator for Deep Brain Stimulation, model 7428. Rx only. Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA. The dual program model 7428 Kinetra Neurostimulator generates electrical signals that are transmitted to the brain. These signals are delivered from the neurostimulator to the brain via DBS extension or DBS leads. The neurostimulator consists of electronic circuitry and a battery, which are hermetically sealed in a titanium case. The operation of the neurostimulator is supported by a clinician programmer, a therapy controller, and a control magnet.

Recall number
Z-0290-2007
Recalling firm
Medtronic Neurological
Product
Medtronic Kinetra Dual Program Neurostimulator for Deep Brain Stimulation, model 7428. Rx only. Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA. The dual program model 7428 Kinetra Neurostimulator generates electrical signals that are transmitted to the brain. These signals are delivered from the neurostimulator to the brain via DBS extension or DBS leads. The neurostimulator consists of electronic circuitry and a battery, which are hermetically sealed in a titanium case. The operation of the neurostimulator is supported by a clinician programmer, a therapy controller, and a control magnet.
Status
Terminated
Initiated
2006-07-19
Posted
2006-12-19
Terminated
2007-04-18
Reason
A subset of Kinetra implantable neurostimulators may experience a failure of wire connections between the electronic hybrid circuit and battery which may lead to sudden cessation of therapy. Sudden cessation of therapy can result in the immediate return or worsening of underlying symptoms due to the progression of the disease state.
Root cause
Other
Action
Letter dated July 2006. Recommendations were made to Health Care Providers of the following: 1.Consider the potential for this anomaly if your patient present with a device in a POR state. 2. Consider informing your patients of this risk and advise them to seek attention immediately if they experience a return of pre-implant or more advanced symptoms. 3. Determine whether device replacement is warranted based upon consultation with your patient, review of the patient''s medical history and consideration of the relative risks of abrupt cessation of stimulation versus the risk of a device replacement surgical procedure. 4. Report any malfunction, removal and replacement to Medtronic and FDA suing the medical device reporting system. 5. Acknowledge this communication using the reply form enclosed.
Quantity
6726 devices
Distribution
Worldwide, including USA, Puerto Rico, Australia, Austria, Belgium, Canada, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Israel, Italy, Jordan, Lebanon, Luxembourg, Netherlands, Norway, Poland, Portugal, Russia, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom.
Lot, serial or model codes
serial numbers in the following ranges:   ¿ NFD100006 - NFD100325H ¿ NFD620017S - NFD625737S ¿ NFD200201K - NFD201303K
510(k) numbers
not on file
PMA numbers
["P960009S027"]
Product code
MHY
Firm FEI
2182207
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register