Blue Ortho: Instructions For Use document of the GPS Trackers.

Recall number
Z-0288-2021
Recalling firm
Blue Ortho
Product
Instructions For Use document of the GPS Trackers.
Status
Terminated
Initiated
2020-09-10
Posted
not on file
Terminated
2023-05-31
Reason
Appropriateness of the language used in the Tracker reprocessing instruction. Reason is that some Low Temperature sterilization conditions mentioned in the IFUs are approved based on rational documents.
Root cause
Other
Action
On 9/10/2020 the firm sent an Urgent Field Safety Notification to its consignees US FDA expressed concern with the appropriateness of the language used in the Tracker reprocessing instruction. Reason is that some Low Temperature sterilization conditions mentioned in the IFUs are approved based on rational documents, leveraged from proper validation files. Based on this concern, Blue Ortho decided to revise the appropriateness of the language used in the Tracker reprocessing instruction LEFUS Tracker for Low Temperature Sterilization v1.0 to remove such conditions. Blue Ortho has already updated the GPS Trackers IFUs, to remove the sterilizations cycle based on rational and not effectively supported by validation. The new version of the IFU is LEFUS Tracker for Low Temperature Sterilization v1.2 (version 1.1 has not been released to the field). The LEFUS Tracker for Low Temperature Sterilization v1.2 is enclosed to this notice. Transmission of this Field Safety Notice: " This notice needs to be passed on all those who need to be aware within your organization or to any organization where the potentially affected devices or documents have been transferred. " Please transfer this notice to other organizations on which this action has an impact. " Please maintain awareness on this notice.
Quantity
3471 units
Distribution
Domestic Distribution Only: CA, CO, SC and WI.
Lot, serial or model codes
Model number: A100003 (Probe) Serial numbers for model A100003: 02-0200-17, 01-0034-19, 02-0097-19, 06-0160-19, 06-0168-19, 11-0083-18, 01-0072-19, 01-0216-14, 02-0091-19, 02-0195-17, 12-0040-15, 02-0192-17, 02-0199-17, 10-0021-12, 08-0028-19, 02-0104-19, 08-0017-19, 05-0380-19, 05-0371-19, 08-0026-19, 08-0003-19, 05-0362-19, 02-0061-19, 08-0064-12, 05-0381-19, 02-0079-19, 02-0050-17, 01-0205-14, 05-0068-18, 04-0288-19, 02-0067-17, 02-0031-17, 02-0064-17, 10-0009-12, 07-0018-14, 02-0008-18, 02-0060-17, 01-0129-19, 02-0044-17, 04-0293-19, 05-0031-14, 05-0365-19, 06-0192-19, 08-0036-18, 10-0024-17, 10-0030-12, 12-0049-19, 12-0050-19, 12-0059-19, 12-0061-19, 06-0221-19, 02-0090-19, 02-0108-19, 06-0187-19, 06-0161-19, 04-0301-19, 01-0120-19.    Model number: A100005 (Diamond T) Serial numbers for model A100005: 08-0080-16, 09-0032-16, 04-0108-13, 08-0088-16, 09-0028-16, 05-0223-13, 12-0123-18, 06-0041-19, 04-0085-19, 06-0438-18, 01-0232-19, 02-0176-19, 05-0185-14, 09-0022-16, 01-0223-19, 04-0167-16, 06-0564-12, 02-0170-19, 05-0057-14, 11-0100-13, 09-0025-16, 02-0120-17, 04-0119-13, 09-0026-16, 10-0085-12, 09-0020-16, 12-0101-18, 11-0074-19, 10-0079-12, 07-0040-19, 10-0167-19, 03-0048-19, 05-0191-14, 10-0082-12, 07-0251-14, 06-0109-17, 10-0175-19, 10-0040-19, 01-0222-19.   Model number: A100006 (Diamond G) Serial numbers for model A100006: 08-0106-16, 08-0093-16, 01-0342-14, 08-0105-16, 08-0107-16, 09-0136-18, 12-0187-18, 05-0137-19, 04-0185-19, 06-0489-18, 05-0129-19, 06-0064-19, 02-0214-19, 06-0168-17, 11-0087-16, 07-0488-19, 05-0077-16, 02-0218-19, 06-0159-17, 01-0288-19, 02-0180-17, 06-0153-17, 05-0386-14, 03-0375-14, 06-0167-17, 06-0130-17, 11-0098-16, 02-0164-17, 05-0396-14, 06-0143-17, 07-0350-19, 10-0225-19, 03-0120-19, 07-0108-17, 01-0341-14, 10-0237-19, 07-0083-17, 10-0232-19, 05-0066-16.  Model number: A100004 (Diamond F) Serial numbers for model A100004: 04-0013-19, 03-0336-14, 03-0083-14, 01-0014-15, 05-0066-19, 05-0154-14, 01-0200-19, 01-0203-19.
510(k) numbers
["K193618"]
PMA numbers
not on file
Product code
OLO
Firm FEI
3008027912
Firm city
Gieres
Firm state
not on file
Firm country
France

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register