Immunetech Inc: ImmuneTech brand MyAllergyTest, Specimen Collection Kit, Part No. 8014, Provider Pack: 10 test (without lancet), Product is manufactured distributed by ImmuneTech, Inc., Foster City, CA; but the non-sterile alcohol prep pads are manufactured by H & P Industries, Inc., Triad Group, Hartland, WI The MyAIIergyTest is for the collection and transport of a capillary blood sample to a clinical laboratory for allergen specific lgE profile testing.

Recall number
Z-0288-2012
Recalling firm
Immunetech Inc
Product
ImmuneTech brand MyAllergyTest, Specimen Collection Kit, Part No. 8014, Provider Pack: 10 test (without lancet), Product is manufactured distributed by ImmuneTech, Inc., Foster City, CA; but the non-sterile alcohol prep pads are manufactured by H & P Industries, Inc., Triad Group, Hartland, WI The MyAIIergyTest is for the collection and transport of a capillary blood sample to a clinical laboratory for allergen specific lgE profile testing.
Status
Terminated
Initiated
2011-11-04
Posted
2011-11-25
Terminated
2012-06-14
Reason
ImmuneTech is recalling certain Specimen Collection Kits that contain Triad brand non-sterile alcohol prep pads which have been recalled by Triad Group, Hartland, WI due to Bacillus cereus, microbial contamination.
Root cause
Material/Component Contamination
Action
The firm, IMMUNETECH, sent an "URGENT PRODUCT RECALL NOTIFICATION" letter dated November 4, 2011, to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to inspect their inventory and quarantine all affected kits; if they have further distributed the affected kits, contact Immunetech to accurately identify and immediately notify their customers and facilitate the return and exchange of all remaining kits. If you have any questions, call (650) 312-1066, Monday-Friday, 9:00am -5:00pm (pacific) or email: customersupport@immunetech.com.
Quantity
310
Distribution
Worldwide distribution.
Lot, serial or model codes
Lot Numbers: F0024, G0006, G0042, G0047, G0054, H0046
510(k) numbers
["K021698","K061979"]
PMA numbers
not on file
Product code
JKA
Firm FEI
3004091547
Firm city
Foster City
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register