Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems (Cleveland) Inc: Respiratory Gating System, Catalog Number: NPTA540, Model #4535 674 26701, for the following affected CT/PET scanning device systems : GEMINI TF CT/PET Systems 16 Slice, GEMINI TF CT/PET System 16 Slice Mobile; GEMINI TF CT/PET System 64 Slice Mobile; and GEMINI TF CT/PET System 64 Slice. (Model Numbers: 16 slice: 4535 679 83931; 64 slice Mobile: 4535 674 47551; and 64 slice: 4535 679 94741.

Recall number
Z-0287-2010
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Respiratory Gating System, Catalog Number: NPTA540, Model #4535 674 26701, for the following affected CT/PET scanning device systems : GEMINI TF CT/PET Systems 16 Slice, GEMINI TF CT/PET System 16 Slice Mobile; GEMINI TF CT/PET System 64 Slice Mobile; and GEMINI TF CT/PET System 64 Slice. (Model Numbers: 16 slice: 4535 679 83931; 64 slice Mobile: 4535 674 47551; and 64 slice: 4535 679 94741.
Status
Terminated
Initiated
2008-11-24
Posted
2009-11-17
Terminated
2010-09-24
Reason
A leak between the Tube Interface and the Outlet Tube of the transducer for the Respiratory Gating System component may result in a failure of the device to produce respiratory correlated images.
Root cause
Process design
Action
Philips Medical Systems Nuclear Medicine division initiated this recall by sending a 'Customer Advisory Notification to their customer accounts on 11/24/2008. This notification informs the customer of the potential for the device equipped with the Pulmonary Gating Option to develop an air leak which could result in the device being unable to produce the desired respiratory correlated images. The notification provides field test instructions and recommends that the customer test the bellows on each day of use in order to assess that the unit is functioning properly. If the bellows device fails the test the recalling firm recommends that the unit not be used and that the recalling firm be contacted via telephone or E-mail. The firm intends to conduct a field correction of the affected units in which their Field Service Engineers will install upgrade kits on the affected devices tentatively scheduled to begin in September, 2009.
Quantity
51 units
Distribution
Worldwide distribution: USA, Puerto Rico, Canada, Belgium, China, Denmark, Egypt, France, Germany, Hong Kong, India, Japan, Netherlands, Pakistan, Philippines, Poland, Saudi Arabia, Singapore, South Korea, Spain, and Sweden.
Lot, serial or model codes
The affected Serial Numbers are: Gemini TF 16 Slice: 7005, 7052, 7087, 7112, 7121, 7138, 7115, 7135, 7006, 7008, 7015, 7021, 7020, 7026, 7035, 7039, 7041, 7049, 7053, 7058, 7060, 7069, 7075, 7080, 7083, 7084, 7098, 7117, 7118, 7142; Gemini TF 16 Slice Mobile: 7059, 7104; Gemini TF 64 Slice Mobile: 7092; Gemini TF 64 Slice: 7012, 7013, 7032, 7071, 7082, 7091, 7126, 7132, 7088, 7103, 7136, 7027, 7109, 7105, 7028, 7167, 7078, and 7110.
510(k) numbers
["K033357","K052640"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register