Randox Laboratories Ltd.: Rx series HbA1c REF/Catalog Number HA3830, GTIN 05055273203561, IVD; -- RANDOX LABORATORIES LTD. , This product is suitable for use on RX series instruments which includes the RX Daytona and RX Imola analysers. Packaging Format: HbA1c R1: Antibody Reagent - 3 x 14 ml HbA1c R2: Agglutinator Reagent - 3 x 14 ml R3: Haemoglobin Denaturant Reagent - 3 x 50 ml Hb R1: Total Haemoglobin Reagent - 3 x 28 ml For Professional Use in the quantitative in vitro determination of Haemoglobin A1c concentration in whole blood.
- Recall number
- Z-0285-2018
- Recalling firm
- Randox Laboratories Ltd.
- Product
- Rx series HbA1c REF/Catalog Number HA3830, GTIN 05055273203561, IVD; -- RANDOX LABORATORIES LTD. , This product is suitable for use on RX series instruments which includes the RX Daytona and RX Imola analysers. Packaging Format: HbA1c R1: Antibody Reagent - 3 x 14 ml HbA1c R2: Agglutinator Reagent - 3 x 14 ml R3: Haemoglobin Denaturant Reagent - 3 x 50 ml Hb R1: Total Haemoglobin Reagent - 3 x 28 ml For Professional Use in the quantitative in vitro determination of Haemoglobin A1c concentration in whole blood.
- Status
- Terminated
- Initiated
- 2017-06-07
- Posted
- not on file
- Terminated
- 2018-07-02
- Reason
- Randox confirmed low calibration absorbance with HbA1c, Catalog Number HA3830, Batch 414553. The reagent may not continue to meet its specific performance characteristics throughout the shelf life of the product.
- Root cause
- Under Investigation by firm
- Action
- Randox Laboratories sent an Urgent Field Safety Notice dated June 6, 2017, sent to all global consignees on June 7, 2017, via e-mail. This includes the sole US consignee. Customers are advised to dispose of any existing stock and to contact technical support with respect to reimbursement for loss of material. The company will follow up every 2 weeks over a six week period with the consignees. For further questions, please call (304) 728-2890.
- Quantity
- 90 kits
- Distribution
- Puerto Rico; International: Portugal, Italy, Turkey, Slovakia, Russia, Thailand and Vietnam
- Lot, serial or model codes
- Lot/Batch 414553, expiry 28th Jan 2019
- 510(k) numbers
- ["K021897"]
- PMA numbers
- not on file
- Product code
- LCP
- Firm FEI
- 1000361607
- Firm city
- Crumlin
- Firm state
- North
- Firm country
- Ireland
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28