Skip to the content

FDA Medical Device Recalls (openFDA)

Randox Laboratories Ltd.: Rx series HbA1c REF/Catalog Number HA3830, GTIN 05055273203561, IVD; -- RANDOX LABORATORIES LTD. , This product is suitable for use on RX series instruments which includes the RX Daytona and RX Imola analysers. Packaging Format: HbA1c R1: Antibody Reagent - 3 x 14 ml HbA1c R2: Agglutinator Reagent - 3 x 14 ml R3: Haemoglobin Denaturant Reagent - 3 x 50 ml Hb R1: Total Haemoglobin Reagent - 3 x 28 ml For Professional Use in the quantitative in vitro determination of Haemoglobin A1c concentration in whole blood.

Recall number
Z-0285-2018
Recalling firm
Randox Laboratories Ltd.
Product
Rx series HbA1c REF/Catalog Number HA3830, GTIN 05055273203561, IVD; -- RANDOX LABORATORIES LTD. , This product is suitable for use on RX series instruments which includes the RX Daytona and RX Imola analysers. Packaging Format: HbA1c R1: Antibody Reagent - 3 x 14 ml HbA1c R2: Agglutinator Reagent - 3 x 14 ml R3: Haemoglobin Denaturant Reagent - 3 x 50 ml Hb R1: Total Haemoglobin Reagent - 3 x 28 ml For Professional Use in the quantitative in vitro determination of Haemoglobin A1c concentration in whole blood.
Status
Terminated
Initiated
2017-06-07
Posted
not on file
Terminated
2018-07-02
Reason
Randox confirmed low calibration absorbance with HbA1c, Catalog Number HA3830, Batch 414553. The reagent may not continue to meet its specific performance characteristics throughout the shelf life of the product.
Root cause
Under Investigation by firm
Action
Randox Laboratories sent an Urgent Field Safety Notice dated June 6, 2017, sent to all global consignees on June 7, 2017, via e-mail. This includes the sole US consignee. Customers are advised to dispose of any existing stock and to contact technical support with respect to reimbursement for loss of material. The company will follow up every 2 weeks over a six week period with the consignees. For further questions, please call (304) 728-2890.
Quantity
90 kits
Distribution
Puerto Rico; International: Portugal, Italy, Turkey, Slovakia, Russia, Thailand and Vietnam
Lot, serial or model codes
Lot/Batch 414553, expiry 28th Jan 2019
510(k) numbers
["K021897"]
PMA numbers
not on file
Product code
LCP
Firm FEI
1000361607
Firm city
Crumlin
Firm state
North
Firm country
Ireland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register