Canon Medical System, USA, INC.: Cartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D, is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.
- Recall number
- Z-0284-2025
- Recalling firm
- Canon Medical System, USA, INC.
- Product
- Cartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D, is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.
- Status
- Open, Classified
- Initiated
- 2024-09-17
- Posted
- 2024-11-01
- Terminated
- not on file
- Reason
- When PET-CT system is executing reconstruction, if PET acquisition for another patient is performed (or PET reconstruction for another patient is performed from raw data processing), PET reconstruction control processing may hang up and PET reconstruction may stop as a result, which may necessitate repeat PET/CT examinations.
- Root cause
- Software design
- Action
- On 9/17/24, correction notices were emailed to customers asking them to do the following: 1) The modified software will be installed to all applicable systems to correct the occurrence of this problem, which should be available 09/2024. 2) To avoid the problem, until the corrective actions have been taken, it is strongly recommended that queue of more than one reconstruction be avoided. 3) If PET reconstruction stops unexpectedly, perform the following steps: <1> Execute "Reset PET System" and reboot the PET system. <2> Delete the remaining jobs in the reconstruction queue. <3> After the PET system has rebooted, start up raw data processing and select the appropriate data. <4> Perform one of the following from the pull-down menu for "Helical for AC". . Specify the helical scan line number to perform CT reconstruction. . Specify "Select Image" and select the CT data used for attenuation correction. <5> Perform PET reconstruction. If PET reconstruction cannot be performed because the CT data for attenuation correction is not saved or for any other reason, follow the procedure below. Specifying helical scan data used for attenuation correction. For helical + PET acquisition, the helical scan data used for attenuation correction can be specified. When "Original" is specified, the CT data used for reconstruction synchronized with scanning is used. When the helical scan line number is specified, reconstruction for helical scan is performed first. When "Select Image" is specified, the data used for attenuation correction can be selected from the displayed list, see Image Reconstruction in the operation manual. 4) Share this information with all users, radiologists, clinical engineering, or biomedical groups. 5) Complete and return the response form via email to RAffairs@us.medical.canon 6) If you have questions please email them to the address above.
- Quantity
- 27
- Distribution
- US: NV, AR, MN, IA, IL, NC, LA, KS, MS, CA, AL, MD, CO, IN, TN, AZ, WI, VA
- Lot, serial or model codes
- UDI-DI: 04987670103810, Software: V10.8 SP0013, V10.15 SP0006. Serial Numbers: 3DA1982001, 3DA2072002, 3DA2072003, 3DA20X2004, 3DA20Y2005, 3DA2112006, 3DA2132007, 3DA2162008, 3DA2192009, 3DA2242010, 3DA2252011, 3DA2262012, 3DA2272013, 3DA2272014, 3DA2282015, 3DA2292016, 3DA22Y2017, 3DA22Z2018, 3DA2322019, 3DA2322020, 3DA2332021, 3DA2332022, 3DA2342023, 3DA2352024, 3DA2352025, 3DB2362026, 3DB2422029
- 510(k) numbers
- ["K231748"]
- PMA numbers
- not on file
- Product code
- KPS
- Firm FEI
- 2020563
- Firm city
- Tustin
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28