FDA Medical Device Recalls (openFDA)
Animas Corporation: One Touch Ping Glucose Monitoring System Combines the functionality of OneTouch(R) Ping (TM) Insulin Pump and a OneTouch(R) Ping (TM) Meter Remote through radio frequency (RF) communication. Together, they provide an option to help make insulin delivery more discreet and flexible. Also, allows the most recent blood glucose results from the OneTouch(R) Ping (TM) Meter Remote to be automatically entered into bolus insulin calculations to cover carbohydrates in food or to correct for a high blood glucose level.
- Recall number
- Z-0283-2009
- Recalling firm
- Animas Corporation
- Product
- One Touch Ping Glucose Monitoring System Combines the functionality of OneTouch(R) Ping (TM) Insulin Pump and a OneTouch(R) Ping (TM) Meter Remote through radio frequency (RF) communication. Together, they provide an option to help make insulin delivery more discreet and flexible. Also, allows the most recent blood glucose results from the OneTouch(R) Ping (TM) Meter Remote to be automatically entered into bolus insulin calculations to cover carbohydrates in food or to correct for a high blood glucose level.
- Status
- Terminated
- Initiated
- 2008-08-04
- Posted
- 2008-11-07
- Terminated
- 2008-11-12
- Reason
- display old, inaccurate values for bolus amount delivered and amount of planned bolus totals
- Root cause
- Software design
- Action
- On 8/6/08, the recalling firm telephoned all users to inform them of problem and the need to return the system.
- Quantity
- 251 units
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- Part numbers: 100-430-00, 100-431-00, 100-432-00, 100-434-00, and 100-435-00. Serial numbers: 01-02000-15, 02-02001-15, 03-02002-15, 04-02003-15, 04-02010-15, 05-02004-15, 05-02011-15, 06-02012-15, 06-02100-15, 07-02013-15, 07-02101-15, 08-02007-15, 08-02102-15, 09-02008-15, 09-02110-15, 10-02009-15, 10-02104-15, 10-02111-15, 11-02112-15, 12-02032-15, 12-02106-15, 12-02113-15, 12-02120-15, 13-02033-15, 13-02040-15, 13-02107-15, 13-02121-15, 13-02202-15, 14-02027-15, 14-02034-15, 14-02108-15, 14-02203-15, 14-02210-15, 15-02028-15, 15-02035-15, 15-02042-15, 15-02123-15, 15-02204-15, 16-02029-15, 16-02036-15, 16-02050-15, 16-02117-15, 16-02124-15, 16-02131-15, 16-02205-15, 16-02300-15, 17-02037-15, 17-02051-15, 17-02118-15, 17-02132-15, 17-02213-15, 17-02220-15, 18-02038-15, 18-02126-15, 18-02140-15, 18-02221-15, 19-02039-15, 19-02053-15, 19-02060-15, 19-02127-15, 19-02208-15, 19-02222-15, 20-02047-15, 20-02054-15, 20-02142-15, 20-02209-15, 20-02216-15, 20-02223-15, 20-02304-15, 21-02048-15, 21-02129-15, 21-02136-15, 21-02143-15, 21-02150-15, 21-02217-15, 21-02224-15, 21-02231-15, 22-02049-15, 22-02056-15, 22-02063-15, 22-02144-15, 22-02151-15, 22-02225-15, 22-02232-15, 23-02057-15, 23-02071-15, 23-02152-15, 23-02219-15, 23-02233-15, 23-02240-15, 24-02058-15, 24-02065-15, 24-02072-15, 24-02139-15, 24-02160-15, 24-02241-15, 25-02059-15, 25-02066-15, 25-02080-15, 25-02147-15, 25-02228-15, 25-02235-15, 26-02148-15, 26-02155-15, 26-02162-15, 26-02236-15, 26-02243-15, 26-02250-15, 26-02331-15, 27-02149-15, 27-02156-15, 27-02163-15, 27-02237-15, 27-02244-15, 28-02076-15, 28-02090-15, 28-02157-15, 28-02164-15, 28-02171-15, 28-02245-15, 28-02319-15, 28-02333-15, 29-02091-15, 29-02158-15, 29-02172-15, 29-02246-15, 29-02253-15, 29-02260-15, 30-02092-15, 30-02166-15, 30-02173-15, 30-02180-15, 30-02247-15, 30-02254-15, 30-02261-15, 30-02342-15, 31-02093-15, 31-02167-15, 31-02248-15, 31-02255-15, 31-02262-15, 31-02336-15, 32-02094-15, 32-02168-15, 32-02175-15, 32-02256-15, 32-02263-15, 32-02351-15, 33-02088-15, 33-02169-15, 33-02176-15, 33-02190-15, 33-02345-15, 34-02096-15, 34-02177-15, 34-02184-15, 34-02191-15, 34-02258-15, 35-02178-15, 35-02185-15, 35-02192-15, 35-02259-15, 36-02098-15, 36-02186-15, 36-02193-15, 36-02348-15, 37-02099-15, 37-02195-15, 39-02189-15, 39-02196-15, 40-01901-15, 40-02197-15, 41-02198-15, 42-01910-15, 42-02199-15, 43-01911-15, 44-01912-15, 45-01906-15, 45-01913-15, 45-01920-15, 48-01909-15, 49-01924-15, 49-01931-15, 50-01918-15, 50-01925-15, 51-01926-15, 51-01940-15, 52-01927-15, 52-01934-15, 52-01941-15, 53-01928-15, 53-01942-15, 54-01936-15, 54-01943-15, 54-01950-15, 55-01951-15, 56-01864-15, 56-01938-15, 56-01952-15, 57-01939-15, 57-01946-15, 57-01953-15, 57-01960-15, 58-01866-15, 58-01873-15, 58-01947-15, 58-01961-15, 59-01874-15, 59-01948-15, 59-01948-15, 60-01868-15, 60-01882-15, 60-01949-15, 60-01963-15, 60-01970-15, 61-01869-15, 61-01876-15, 61-01883-15, 61-01890-15, 61-01957-15, 62-01877-15, 62-01884-15, 62-01891-15, 62-01958-15, 63-01885-15, 63-01892-15, 63-01959-15, 63-01973-15, 64-01886-15, 64-01893-15, 64-01974-15, 64-01981-15, 65-01887-15, 67-01889-15, 67-01896-15, 67-01991-15, 68-01897-15, 68-01985-15, 68-01992-15, 69-01898-15, 69-01979-15, 69-01993-15, 70-01899-15, 70-01987-15, 70-01994-15, 71-01995-15, 72-01996-15, 73-01997-15, 74-01998-15, and 75-01999-15.
- 510(k) numbers
- ["K080639"]
- PMA numbers
- not on file
- Product code
- LZG
- Firm FEI
- 3002920072
- Firm city
- West Chester
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28