Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF): Medtronic CareLink SmartSync Device Manager application software (D00U005) used by the following Medtronic Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds): Cobalt XT VR: DVPA2D1, DVPA2D4; Cobalt VR: DVPB3D1, DVPB3D4; Cobalt XT DR: DDPA2D1, DPA2D4; Cobalt DR: DDPB3D1, DDPB3D4; Cobalt XT HF: DTPA2D4, DTPA2D1; Cobalt XT HF; Quad: DTPA2QQ, DTPA2Q1; Cobalt HF: DTPB2D4, DTPB2D1; Cobalt HF Quad: DTPB2QQ, DTPB2Q1. Crome VR: DVPC3D1, DVPC3D4; Crome DR: DDPC3D1, DDPC3D4; Crome HF: DTPC2D4, DTPC2D1; Crome HF Quad: DTPC2QQ, DTPC2Q1.

Recall number
Z-0282-2022
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Product
Medtronic CareLink SmartSync Device Manager application software (D00U005) used by the following Medtronic Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds): Cobalt XT VR: DVPA2D1, DVPA2D4; Cobalt VR: DVPB3D1, DVPB3D4; Cobalt XT DR: DDPA2D1, DPA2D4; Cobalt DR: DDPB3D1, DDPB3D4; Cobalt XT HF: DTPA2D4, DTPA2D1; Cobalt XT HF; Quad: DTPA2QQ, DTPA2Q1; Cobalt HF: DTPB2D4, DTPB2D1; Cobalt HF Quad: DTPB2QQ, DTPB2Q1. Crome VR: DVPC3D1, DVPC3D4; Crome DR: DDPC3D1, DDPC3D4; Crome HF: DTPC2D4, DTPC2D1; Crome HF Quad: DTPC2QQ, DTPC2Q1.
Status
Open, Classified
Initiated
2021-10-12
Posted
not on file
Terminated
not on file
Reason
The processing of collected episode data may fail due to a software error.
Root cause
Software design
Action
The recalling firm began issuing the correction notification dated October 2021 on 10/12/2021 via 2-day UPS delivery that was addressed to risk managers or clinical managers. The letter recommends physicians follow normal clinical practices given the implanted Cobalt/Crome devices will continue to operate as programmed. If a failure to interrogate a Cobalt or Crome device occurs with a SmartSync programmer, the physician is to confirm that the SmartSync application software has been updated to D00U005 version 5.0.0 (or higher). They are to contact their Medtronic representative or Tachy Technical Services at 800-723-4636 for assistance with retrieving the session data. If a CareLink transmission is attempted, but the transmission is not viewable on the CareLink network (i.e., the transmission is missing from the transmission list for the patient), the physician is to contact Medtronic Technical Services at 800-723-4636 for assistance. The notice is to be shared with those who need to be aware within the consignee's organization or with any organization where the SmartSync programmers may have been transferred. Consignees are asked to sign and return the enclosed confirmation form and confirm that the SmartSync tablets have been updated to Cobalt/Crome application software D00U005, version 5.0.0 or higher.
Quantity
18,715 devices
Distribution
Worldwide distribution - US Nationwide. There was military/government distribution. The countries of Andorra, Australia, Austria, Belgium, Brunei Darussalam, Canada, Czech Republic, Denmark, Finland, France, French Guiana, French Polynesia, Germany, Greece, Guadeloupe, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Kuwait, Liechtenstein, Luxembourg, Martinique, Netherlands, New Caledonia, New Zealand, Norway, Poland, Portugal, Puerto Rico, Reunion, Romania, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
Lot, serial or model codes
Application software lower than v5.0.0.
510(k) numbers
not on file
PMA numbers
["P010031S750"]
Product code
OSR
Firm FEI
2182208
Firm city
Mounds View
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register