GE Healthcare, LLC: Giraffe OmniBed Carestation CS1 - Product Usage: The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo regulate based on their own physiology.

Recall number
Z-0281-2020
Recalling firm
GE Healthcare, LLC
Product
Giraffe OmniBed Carestation CS1 - Product Usage: The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo regulate based on their own physiology.
Status
Open, Classified
Initiated
2019-07-26
Posted
not on file
Terminated
not on file
Reason
Fasteners potentially torqued to a value less than the specified value and can potentially cause the canopy to become detached and fall.
Root cause
Nonconforming Material/Component
Action
GE Healthcare notified customers on about 07/26/2019 via phone calls and a "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were instructed to immediately stop use of and segregate the affected devices, that GE Healthcare would correct all affected products at no cost, and a GE Healthcare representative will contact them to arrange for the correction. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or your local Service Representative. A second customer letter was sent on 08/05/2019 to all customer. The second letter contained the same instructions, but provided an updated list of affected serial numbers.
Quantity
59 devices (50 US, 9 OUS)
Distribution
Worldwide distributions - US Nationwide in the states of CA, CT, FL, ID, IL, KS, MI, NC, NC, NJ, NM, PA, TX, VA. Countries of China, France, Germany, Korea, Republic of China.
Lot, serial or model codes
Model Number: 2082844-001;   Serial Number (UDI Number): TABY70690 (010084068211686221TABY7069011190610), TABY70692 (010084068211686221TABY7069211190611), TABY70708 (010084068211686221TABY7070811190612), TABY70719 (010084068211686221TABY7071911190613), TABY70709 (010084068211686221TABY7070911190612), TABY70724 (010084068211686221TABY7072411190614), TABY70722 (010084068211686221TABY7072211190614), TABY70718 (010084068211686221TABY7064211190603), TABY70717 (010084068211686221TABY7064411190603), TABY70712 (010084068211686221TABY7064511190603), TABY70678 (010084068211686221TABY7067811190607), TABY70725 (010084068211686221TABY7072511190615), TABY70672 (010084068211686221TABY7067211190607), TABY70686 (010084068211686221TABY7068611190610), TABY70688 (010084068211686221TABY7068811190610), TABY70697 (010084068211686221TABY7069711190611), TABY70726 (010084068211686221TABY7072611190616), TABY70729 (010084068211686221TABY7072911190616), TABY70645 (010084068211686221TABY7064711190604), TABY70647 (010084068211686221TABY7066011190605), TABY70646 (010084068211686221TABY7071211190613), TABY70644 (010084068211686221TABY7071711190613), TABY70642 (010084068211686221TABY7071811190613), TABY70732 (010084068211686221TABY7073211190617), TABY70723 (010084068211686221TABY7072311190614), TABY70747 (010084068211686221TABY7074711190619), TABY70748 (010084068211686221TABY7074811190619), TABY70673 (010084068211686221TABY7067311190607), TABY70676 (010084068211686221TABY7067611190607), TABY70693 (010084068211686221TABY7069311190611), TABY70648 (010084068211686221TABY7064811190604), TABY70675 (010084068211686221TABY7067511190607), TABY70694 (010084068211686221TABY7069411190611), TABY70695 (010084068211686221TABY7069511190611), TABY70730 (010084068211686221TABY7073011190617), TABY70734 (010084068211686221TABY7073411190617), TABY70737 (010084068211686221TABY7073711190618), TABY70677 (010084068211686221TABY7067711190607), TABY70698 (010084068211686221TABY7069811190611), TABY70699 (010084068211686221TABY7069911190611), TABY70703 (010084068211686221TABY7070311190612), TABY70706 (010084068211686221TABY7070611190612), TABY70727 (010084068211686221TABY7072711190616), TABY70728 (010084068211686221TABY7072811190616), TABY70660 (010084068211686221TABY7064611190603), TABY70689 (010084068211686221TABY7068911190610), TABY70705 (010084068211686221TABY7070511190612), TABY70707 (010084068211686221TABY7070711190612), TABY70731 (010084068211686221TABY7073111190617), TABY70733 (010084068211686221TABY7073311190617), TABY70652 (Not Available),  TABY70654 (Not Available), TABY70662 (Not Available), TABY70664 (Not Available), TABY70710 (Not Available), TABY70715 (Not Available), TABY70716 (Not Available), TABY70753 (Not Available), TABY70659 (Not Available), TABY70704 (010084068211686221TABY7070411190612),  TABY70755 (010084068211686221TABY7075511190620), TABY70638 (Not Available), TABY70649 (Not Available), TABY70656 (Not Available), TABY70661 (Not Available), TABY70657 (Not Available), TABY70650 (Not Available), TABY70667 (Not Available), TABY70668 (Not Available), TABY70653 (Not Available), TABY70663 (Not Available), TABY70670 (Not Available), TABY70665 (Not Available), TABY70669 (Not Available), TABY70671 (Not Available), TABY70714 (Not Available), TABY70658 (Not Available), TABY70655 (Not Available), TABY70666 (Not Available), TABY70651 (Not Available).
510(k) numbers
["K152814"]
PMA numbers
not on file
Product code
FMZ
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register