Siemens Medical Solutions USA, Inc: Si emans Simiae 3000 Cassette Holder Electronic Imaging device (Image Identifier), with CT Part No: 19 57-161 or without CT Part No. 19 57 153; Siemens Medical Solutions USA, Concord, CA 94520

Recall number
Z-0281-2008
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Si emans Simiae 3000 Cassette Holder Electronic Imaging device (Image Identifier), with CT Part No: 19 57-161 or without CT Part No. 19 57 153; Siemens Medical Solutions USA, Concord, CA 94520
Status
Terminated
Initiated
2007-02-27
Posted
2008-01-31
Terminated
2011-02-03
Reason
Loose Mounting Screws -- Cassette holder may become loose and result in the device dislodging and colliding with the patient.
Root cause
Labeling design
Action
Customers were sent a Customer Safety Advisory Notice on February 27, 2007, notifying them that the mounting screws on the cassette holders on the SIMVIEW 3000 and NT Simulators, as well as some MEVASIM Simulators, could loosen and can cause damage to the system, or personal injury. The firm requested that they check of tightness on the mounting hardware. A STOP Modification notice was issued on 06/01/2007, to all affected regional service organizations that are responsible for implementing updates due to new information. Updated letters and User instructions will be released in August 2007.
Quantity
287 units total, all recalls Z-0281-285-2008
Distribution
Worldwide Distribution - USA including states of IN, VA, MA, VT, NJ, NY, PA, DC, NC, SC, GA, AL, TN, MI, KY, OH, IN MI, MN, ND, MT, IL, MO, KS, NE, AR, OK, TX, AZ, CA, SD, LA, Puerto Rico, and countries of Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Colombia, Costa Rica, Croatia, Denmark, Ecuador, Egypt, El Salvador, Estonia, France, Germany, Greece, Hungary, India, Iran, Ireland, Italy, Kuwait, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, China, Israel/Palestine, Philippines, Poland, Portugal, Qatar, Republic of Korea, Russia, Singapore, South Africa, Spain, Thailand, Turkey, Ukraine, UK, and Vietnam.
Lot, serial or model codes
Serial Numbers: 4164, 4062, 4114, 4037, 4155, 4125, 4180, 4096, 4064, 4107, 4145, 4144, 4143, 4081, 4172, 4152, 4090, 4161, 4181, 4093, 4066, 4054, 4091, 4068, 4128, 4179, 4019, 4023, 4117, 4094, 4015, 4083, 4174, 4074, 4082, 4011, 4078, 4034, 4063, 4025, 4171, 4086 4134, 4029, 4002, 4003, 4004, 4027, 4053, 4061, 4118, 4123, 4124, 4008, 4009, 4010, 4014, 4016, 4017, 4021, 4022, 4024, 4026, 4031, 4032, 4035, 4043, 4059, 4072, 4075, 4076, 4078, 4085, 4087, 4092, 4095, 4098, 4102, 4104, 4105, 4108, 4109, 4110, 4111, 4112, 4115, 4116, 4130, 4136, 4142, 4150, 4153, 4156, 4173, 70001004511, 40000296842, 4099, 4147, 4100, 4154, 4044, 4119, 4121, 4122, 4131, 4139, 4185, 4020, 4030, 4036, 4079, 4135, 4167, 4077, 4039, 4071, 4120, 41514127, 4176, 4097, 4049, 4046, 4070, 4084, 4138, 4160, 4165, 4028, 4042, 4101, 410, 4158, 4045, 4065, 4060, 4088, 4177, 4073, 4113, 4175, 4141, 4040, 4137, and 4052
510(k) numbers
["K946053"]
PMA numbers
not on file
Product code
KPQ
Firm FEI
2937457
Firm city
Concord
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register