Skip to the content

FDA Medical Device Recalls (openFDA)

Cyberonics, Inc: Cyberonics Vagus Nerve Stimulator (VNS) Therapy AspireSR Generator, Model 106

Recall number
Z-0280-2016
Recalling firm
Cyberonics, Inc
Product
Cyberonics Vagus Nerve Stimulator (VNS) Therapy AspireSR Generator, Model 106
Status
Terminated
Initiated
2015-10-07
Posted
2015-11-17
Terminated
2017-09-06
Reason
Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit potential for decreased battery longevity.
Root cause
Manufacturing material removal
Action
Cyberonics sent written recall notifications beginning October 14, 2015, to all affected customers. The letter identified the product the problem and the action to be taken by the customer. The notice also directs customers to contact Cyberonics to schedule additional on-site troubleshooting. The on-site troubleshooting proposed by Cyberonics will utilize a feature included within the Model 250 Version 9.0 Programming Software as an alternate method of verifying heart rate and determining that the appropriate heart beat sensitivity setting is programmed. The Version 9.0 software displays an instantaneous heart rate when the user selects the 'Refresh' feature. This heart rate is measured directly by the generator and displayed on the programmer. For further questions, please call (866) 882-8804.
Quantity
27 units
Distribution
US Distribution to the states of : OR, NY, GA, PA, CA, NC, KS, FL, KY, MI, TX, IL, OH, IN, UT, CO and WV.
Lot, serial or model codes
Serial Numbers: 37719, 37899, 38123, 38414, 38428, 39385, 39531, 39607, 39801, 39821, 40037, 40048, 40249, 40251, 40271, 40354, 40428, 40488, 40526, 40603, 40751, 41000, 41109, 41152, 41805, 42022, 42780
510(k) numbers
not on file
PMA numbers
["P970003"]
Product code
LYJ
Firm FEI
1000117004
Firm city
Houston
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register