Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific: Stainless Steel Greenfield Vena Cava Filter with 12 Fr / 4,0 mm FlexCarrier Capsule. For Femoral Vein Introduction Only. Order No. 50-501. Sterile EO. Single Use Only. Read Instructions for Use before using this device. Distributed in USA by: Boston Scientific Corporation, One Boston Scientific Place, Natick, MA USA 01760. Manufacturer: Boston Scientific Cork Ltd., Business & Technology Park, Model Farm Road, Cork, Ireland. The Stainless Steel Greenfield Vena Cava Filter with 12 Fr. / 4,0 mm introducer System is a permanently implanted stainless steel device designed to protect against pulmonary embolism while maintaining patency of the inferior vena cava. The Stainless Steel Greenfield Vena Cava Filter comes preloaded in a 12 Fr. / 4,0 mm jugular or femoral introducer catheter.

Recall number
Z-0280-06
Recalling firm
Boston Scientific
Product
Stainless Steel Greenfield Vena Cava Filter with 12 Fr / 4,0 mm FlexCarrier Capsule. For Femoral Vein Introduction Only. Order No. 50-501. Sterile EO. Single Use Only. Read Instructions for Use before using this device. Distributed in USA by: Boston Scientific Corporation, One Boston Scientific Place, Natick, MA USA 01760. Manufacturer: Boston Scientific Cork Ltd., Business & Technology Park, Model Farm Road, Cork, Ireland. The Stainless Steel Greenfield Vena Cava Filter with 12 Fr. / 4,0 mm introducer System is a permanently implanted stainless steel device designed to protect against pulmonary embolism while maintaining patency of the inferior vena cava. The Stainless Steel Greenfield Vena Cava Filter comes preloaded in a 12 Fr. / 4,0 mm jugular or femoral introducer catheter.
Status
Terminated
Initiated
2005-12-02
Posted
2005-12-20
Terminated
2007-04-02
Reason
There have been reports of detachment at the bond between the carrier capsule and the outer sheath of the Greenfield Vena Cava Filters with 12 Fr Femoral introducer Systems manufactured before Marcyh 10, 2004. If the capsule should detach during an implantation procedure, there is a risk of cardiac and pulmonary embolization.
Root cause
Other
Action
Recall packages started going out 12/02/05 via Fed Ex and was completed 12/03/05. A letter was directed to several persons within each hospital . The letter describes the issues, the affected lot numbers and the complaints that Boston Scientific has received. Affected devices are to be segregated for return back to Boston Scientific. A 'Reply Verification Tracking Form' is asked to be returned back to Boston Scientific.
Quantity
18,000
Distribution
nationwide (except ND); Virgin Islands, OUS to include: united Arab Emirates, Argentina, Australia, Belgium, Brunei Darussalam, Brazil, Canada, Chile, China, Colombia, Germany, Spain, France, Ireland, Great Britain, Israel, India, Italy, Jamaica, Jordan, Japan, Korea (South Republic), Kuwait, Lebanon, Mexico, Malaysia, Netherlands, New Zealand, Panama, Philippines, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Thailand, Trinidad and Tobago, Taiwan, Venezuela
Lot, serial or model codes
all codes of product manufactured before March 10, 2004. Catalog no. 50-501 with lot/batch # between 5145758 and 6387904
510(k) numbers
["K964284"]
PMA numbers
not on file
Product code
DTK
Firm FEI
3002095335
Firm city
Maple Grove
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register