Skip to the content

FDA Medical Device Recalls (openFDA)

Accriva Diagnostics, Inc.: AVOXimeter 1000E, Product Number: AVOX1000E. The AVOXimeter 1000E is a battery-operated bedside whole blood oximeter.

Recall number
Z-0279-2025
Recalling firm
Accriva Diagnostics, Inc.
Product
AVOXimeter 1000E, Product Number: AVOX1000E. The AVOXimeter 1000E is a battery-operated bedside whole blood oximeter.
Status
Completed
Initiated
2024-09-27
Posted
2024-10-31
Terminated
not on file
Reason
Values used to calibrate whole blood oximeter instruments were observed to be higher than normal, which may result in incorrect patient results where oxygen levels are higher than actual results, which may lead to patient care mismanagement requiring medical intervention.
Root cause
Process change control
Action
On 9/27/24, a recall notice was sent to a customer informing them to do the following: 1) This recall notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where the potentially affected device has been transferred.. 2. Call Technical Support 1-800-678-0710 to initiate a Service Call and Returned Material Authorization 3. Complete and return the customer response form via email to For technical questions contact Technical Support at the number above or email techsupport.na@werfen.com.
Quantity
1
Distribution
US: AZ
Lot, serial or model codes
UDI-DI: 10711234530023. Serial Number: 6824
510(k) numbers
["K922075"]
PMA numbers
not on file
Product code
DQA
Firm FEI
2250033
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register