FDA Medical Device Recalls (openFDA)
Stryker Neurovascular: Trevo(TM) XP ProVue Procedure Pack, TREVO(TM) XP PROVUE RETRIEVER, 6 mm, 25 mm, .Excelsior(TM) XT-27(TM) 1 TIP STRAIGHT, 150 cm, 6 cm, Microcatheter; UPN 93068 Stroke intervention kit
- Recall number
- Z-0277-2018
- Recalling firm
- Stryker Neurovascular
- Product
- Trevo(TM) XP ProVue Procedure Pack, TREVO(TM) XP PROVUE RETRIEVER, 6 mm, 25 mm, .Excelsior(TM) XT-27(TM) 1 TIP STRAIGHT, 150 cm, 6 cm, Microcatheter; UPN 93068 Stroke intervention kit
- Status
- Terminated
- Initiated
- 2017-11-03
- Posted
- 2017-12-05
- Terminated
- 2019-11-04
- Reason
- Stryker Neurovascular has become aware that some 3-Pack Stroke Fast Packs and Trevo Procedure Packs were manufactured using a carton sleeve where the sleeve label contents did not match the physical contents within the pack.
- Root cause
- Labeling Change Control
- Action
- The firm, Stryker Neurovascular, sent an "Urgent Medical Device Voluntary Recall Immediate Action Required" letter initiating their recall on 11/01/2017. The letter described the product, problem and actions to be taken. The consignees were instructed to immediately check internal inventory; remove and discard the procedure pack carton sleeve; circulate notice; maintain awareness of notice internally until all required actions have been completed with your facility; inform Stryker of any subject devices distributed to other organization, and complete and return customer response form to your nominated Stryker Representative or to NVFieldActions@stryker.com. If you have any questions, call 510-413-2593; email: geraldine.ahern@stryker.com or 510-413-2900.
- Quantity
- 1 unit
- Distribution
- International Distribution to: Germany, Slovakia and Israel.
- Lot, serial or model codes
- Lot Number QX10100894. Exp. Date 28-Oct-18
- 510(k) numbers
- ["K113260","K113778","K132641","K143077","K151667"]
- PMA numbers
- not on file
- Product code
- NRY
- Firm FEI
- 3008853977
- Firm city
- Fremont
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28