Skip to the content

FDA Medical Device Recalls (openFDA)

Synthes USA HQ, Inc.: Synthes External Fixation System Tube to Tube Clamp, Part Number 390.007. Intended for use to provide treatment for long bone and pelvic fractures that require external fixation. Specifically, the components can be used for: Stabilization of soft tissues and fractures; Polytrauma/multiple orthopedic trauma; Vertically stable pelvic fractures, or as treatment adjunct for vertically unstable pelvic fractures; Arthrodeses and osteotomies with soft tissue problems - failures of total joints; Neutralization of fractures stabilized with limited internal fixation; Non-unions/septic non-unions; Intra-operative reductions/stabilization tool to assist with indirect reduction; and Unilateral rectilinear bone segment transport or leg lengthening.

Recall number
Z-0277-2013
Recalling firm
Synthes USA HQ, Inc.
Product
Synthes External Fixation System Tube to Tube Clamp, Part Number 390.007. Intended for use to provide treatment for long bone and pelvic fractures that require external fixation. Specifically, the components can be used for: Stabilization of soft tissues and fractures; Polytrauma/multiple orthopedic trauma; Vertically stable pelvic fractures, or as treatment adjunct for vertically unstable pelvic fractures; Arthrodeses and osteotomies with soft tissue problems - failures of total joints; Neutralization of fractures stabilized with limited internal fixation; Non-unions/septic non-unions; Intra-operative reductions/stabilization tool to assist with indirect reduction; and Unilateral rectilinear bone segment transport or leg lengthening.
Status
Terminated
Initiated
2012-09-28
Posted
2012-11-13
Terminated
2015-09-21
Reason
The insert provided with the product for the specific lot number indicates that this product is MR safe. This product is MR Conditional and is etched as MR Conditional. The insert is incorrect.
Root cause
Labeling mix-ups
Action
Synthes sent an "URGENT NOTICE: MEDICAL DEVICE RECALL" letter dated September 28, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers were asked to complete and return a Verification Form via fax to 610-251-9005. Contact the firm at 610-719-5450 for questions regarding this notice.
Quantity
8
Distribution
Nationwide Distribution-including the states of CA, MN, NC, and TN.
Lot, serial or model codes
Part number 390.007, Lot number 6800599
510(k) numbers
["K082650"]
PMA numbers
not on file
Product code
KTT
Firm FEI
3005180112
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register