OrthoPediatrics Corp: 3.5MM LOCKING SCREW, T15 HEXALOBE, SELF TAPPING, 34MM The OrthoPediatrics Proximal Femur Plating System combines implants and instruments in one convenient system. This system offers the advantages of the osteotomy blade plates and cannulated instrumentation. Osteotomy plates provide ease of reduction and good rotational stability while maintaining bone stock. Cannulated instruments work over a guide wire for precise placement and safety. The subject device is a Bone Screw designed for use with this system.

Recall number
Z-0276-2014
Recalling firm
OrthoPediatrics Corp
Product
3.5MM LOCKING SCREW, T15 HEXALOBE, SELF TAPPING, 34MM The OrthoPediatrics Proximal Femur Plating System combines implants and instruments in one convenient system. This system offers the advantages of the osteotomy blade plates and cannulated instrumentation. Osteotomy plates provide ease of reduction and good rotational stability while maintaining bone stock. Cannulated instruments work over a guide wire for precise placement and safety. The subject device is a Bone Screw designed for use with this system.
Status
Terminated
Initiated
2013-09-26
Posted
2013-11-14
Terminated
2013-12-26
Reason
The firm identified, through complaints that the 3.5MM LOCKING SCREW packaging material was marked as a 3.5mm screw and the part number was correct but that the description indicated a 7.0mm screw. Upon investigation, it has been determined that a lot number 009U2 was packaged and mislabeled internally by OrthoPediatrics.
Root cause
Labeling mix-ups
Action
OrthoPediatrics sent an Urgent Medical Device Recall letter dated September 26, 2013, to all affected customers.The letter identified the product the problem and the actions needed to be taken by the customer The sales representatives and distributors who have been identified via shipping records as having received one or more discrepant devices affected by this Recall were requested to quarantine the devices identified in this Recall by telephone on 18-September-2013. If you or any of your customers have any of these devices in the original packaging please take possession of them and quarantine them in preparation for returning them to OrthoPedlatrics. They are misbranded. If the Screw has already been removed from the original packaging and the original packaging has been discarded and the Screw placed in a sterilization tray, it need not be returned. For further questions please call ( 574) 268-6379.
Quantity
7 units shipped
Distribution
Worldwide Distribution - US including Michigan and the countries of Ireland and Australia
Lot, serial or model codes
Part Number:. 00-0903-2634 and lot number 009U2
510(k) numbers
not on file
PMA numbers
not on file
Product code
HRS
Firm FEI
3006460162
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register