Barcoview: Barco Surgical Display, Model MDSC-2124 The 24-inch MDSC-2124 is a near-patient surgical display that uses widescreen (16:10) LCD technology featuring full High Definition resolution (1920 x 1200). Featuring broad input connectivity, the MDSC-2124 offers a versatile display solution for HD endoscopy cameras, room and boom cameras, ultrasound, PACS and patient information.
- Recall number
- Z-0276-2009
- Recalling firm
- Barcoview
- Product
- Barco Surgical Display, Model MDSC-2124 The 24-inch MDSC-2124 is a near-patient surgical display that uses widescreen (16:10) LCD technology featuring full High Definition resolution (1920 x 1200). Featuring broad input connectivity, the MDSC-2124 offers a versatile display solution for HD endoscopy cameras, room and boom cameras, ultrasound, PACS and patient information.
- Status
- Terminated
- Initiated
- 2008-09-09
- Posted
- 2008-11-07
- Terminated
- 2009-06-02
- Reason
- The front protective cover may loosen and completely fall off.
- Root cause
- Other
- Action
- Consignees were notified via a Field Safety Notice letter on 09/09/2008. They were instructed to inspect the units to check if the front panel is tightly secured to the display housing. Any front panel found to be lose or showing a gap between the front plate and the back shell of the monitor is to be reported the BARCO Service & Support organization and is to be secured using adhesive tape. Follow up Field Safety Notice was sent to consignees on 09/26/2008 indicating the root cause and the firm's corrective field actions implemented October 15, 2008.
- Quantity
- 25 units
- Distribution
- Nationwide Distribution --- including states of AK, GA, IL, and MA.
- Lot, serial or model codes
- All lots
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- RHF
- Firm FEI
- 3004455125
- Firm city
- Kortrijk
- Firm state
- not on file
- Firm country
- Belgium
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28