Skip to the content

FDA Medical Device Recalls (openFDA)

Barcoview: Barco Surgical Display, Model MDSC-2124 The 24-inch MDSC-2124 is a near-patient surgical display that uses widescreen (16:10) LCD technology featuring full High Definition resolution (1920 x 1200). Featuring broad input connectivity, the MDSC-2124 offers a versatile display solution for HD endoscopy cameras, room and boom cameras, ultrasound, PACS and patient information.

Recall number
Z-0276-2009
Recalling firm
Barcoview
Product
Barco Surgical Display, Model MDSC-2124 The 24-inch MDSC-2124 is a near-patient surgical display that uses widescreen (16:10) LCD technology featuring full High Definition resolution (1920 x 1200). Featuring broad input connectivity, the MDSC-2124 offers a versatile display solution for HD endoscopy cameras, room and boom cameras, ultrasound, PACS and patient information.
Status
Terminated
Initiated
2008-09-09
Posted
2008-11-07
Terminated
2009-06-02
Reason
The front protective cover may loosen and completely fall off.
Root cause
Other
Action
Consignees were notified via a Field Safety Notice letter on 09/09/2008. They were instructed to inspect the units to check if the front panel is tightly secured to the display housing. Any front panel found to be lose or showing a gap between the front plate and the back shell of the monitor is to be reported the BARCO Service & Support organization and is to be secured using adhesive tape. Follow up Field Safety Notice was sent to consignees on 09/26/2008 indicating the root cause and the firm's corrective field actions implemented October 15, 2008.
Quantity
25 units
Distribution
Nationwide Distribution --- including states of AK, GA, IL, and MA.
Lot, serial or model codes
All lots
510(k) numbers
not on file
PMA numbers
not on file
Product code
RHF
Firm FEI
3004455125
Firm city
Kortrijk
Firm state
not on file
Firm country
Belgium

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28