Abbott Medical Optics Inc (AMO): Sensar Foldable IOL with OptiEdge. Model number AR40e, 6.0 Diopter. The Sensar Foldable IOLs with OptiEdge Design are ultraviolet-absorbing posterior chamber intraocular lenses and are indicated for the visual correction of aphakia in adult patients in whom a cataractous lens has been removed by extracapsular cataract extraction. The lens are intended to be placed in a capsular bag.

Recall number
Z-0275-2012
Recalling firm
Abbott Medical Optics Inc (AMO)
Product
Sensar Foldable IOL with OptiEdge. Model number AR40e, 6.0 Diopter. The Sensar Foldable IOLs with OptiEdge Design are ultraviolet-absorbing posterior chamber intraocular lenses and are indicated for the visual correction of aphakia in adult patients in whom a cataractous lens has been removed by extracapsular cataract extraction. The lens are intended to be placed in a capsular bag.
Status
Terminated
Initiated
2011-09-20
Posted
2011-11-23
Terminated
2014-01-08
Reason
The firm is initiating this voluntary recall because of the possibility that certain Sensar IOLs are labeled with an incorrect diopter power.
Root cause
Employee error
Action
Abbott Medical Optics, Inc. sent an URGENT FIELD SAFETY NOTICE dated September 21, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to stop using and remove the affected product from their inventory, complete and return the included facsimile form to Abbott Medical Optics and indicate the status of each affected serial number on the form. Customers were asked to return the form even if they have no inventory affected by the recall. For any questions customers were instructed to contact their Abbott Medical Optics Customer Service Representative. For any questions regarding this recall call 714-247-8656.
Quantity
83
Distribution
Worldwide Distribution including China, India, Indonesia, Korea, Malasyia, Taiwan, Thailand, Austria, Finland, France, Germany, Great Britain, Hungary, Italy, Netherlands, Portugal, and Saudy Arabia
Lot, serial or model codes
5185681009 5186101009 5185691009 5186111009 5185701009 5186121009 5185711009 5186131009 5185721009 5186141009 5185731009 5186151009 5185741009 5186161009 5185751009 5186171009 5185761009 5186181009 5185771009 5186191009 5185781009 5186201009 5185791009 5186211009 5185801009 5186221009 5185811009 5186231009 5185821009 5186241009 5185831009 5186251009 5185841009 5186261009 5185851009 5186271009 5185861009 5186281009 5185871009 5186291009 5185881009 5186301009 5185891009 5186311009 5185901009 5186321009 5185911009 5186331009 5185921009 5186341009 5185931009 5186351009 5185941009 5186361009 5185951009 5186371009 5185961009 5186381009 5185971009 5186391009 5185981009 5186401009 5185991009 5186411009 5186001009 5186421009 5186011009 5186431009 5186021009 5186441009 5186031009 5186451009 5186041009 5186461009 5186051009 5186471009 5186061009 5186491009 5186071009 5186501009 5186081009 5186511009 5186091009
510(k) numbers
not on file
PMA numbers
["P980040"]
Product code
HQL
Firm FEI
3003843509
Firm city
Santa Ana
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register