Skip to the content

FDA Medical Device Recalls (openFDA)

Bard Peripheral Vascular Inc: Bard Magnum Disposable Biopsy Core Needle 16 gauge, Catalog Number: MN1613. Bard Magnum Disposable Biopsy Core Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, breast, spleen, lymph nodes and various soft tissue tumors.

Recall number
Z-0274-2010
Recalling firm
Bard Peripheral Vascular Inc
Product
Bard Magnum Disposable Biopsy Core Needle 16 gauge, Catalog Number: MN1613. Bard Magnum Disposable Biopsy Core Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, breast, spleen, lymph nodes and various soft tissue tumors.
Status
Terminated
Initiated
2009-10-08
Posted
2009-11-17
Terminated
2012-04-02
Reason
The recall was initiated after the firm determined that some Magnum Disposable Biopsy Core Needle 16 gauge lots may exhibit an exposed sample chamber, due to extended stylet length resulting from a loose hub connection. The exposed sample chamber will typically result in an inability to obtain the biopsy sample. In the majority of reported cases, there was no injury to the patient or user. When us
Root cause
Process design
Action
The recall notification was initiated by the firm issuing an urgent medical device notification to it's consignees on October 08, 2009 via USPS requesting them to check all inventory locations for the affected products, remove and return affected products to Bard. The consignees were instructed to complete and return a recall effectiveness check form. Bard provided this 800 number 1-800-321-4254 for any questions consignees may have.
Quantity
32, 427 units
Distribution
Nationwide and internationally to Canada Mexico Algeria Andorra Austria Bahrain Belarus Belgium Croatia Cyprus Czech Republic Denmark Finland France French Guiana Germany Gibraltar Greece Hungary Ireland Israel Italy Ivory Coast Kenya Kuwait Latvia Lebanon Luxembourg Malta Martinique Mauritius Morocco Netherlands New Caledonia Nigeria Norway Oman Poland Portugal Republic of Kosovo Republic of Macedonia Reunion Romania Russia Saudi Arabia Senegal Serbia Seychelles Slovenia South Africa Spain Sweden Switzerland Syria Tunisia Turkey Ukraine United Arab Emirates United Kingdom Australia New Zealand Canada Japan Bangladesh China Hong Kong India Indonesia Korea Malaysia Pakistan Singapore Sri Lanka Taiwan Thailand Vietnam Argentina Bahamas Barbados Brazil Chile Colombia Dominican Republic Mexico Peru and Suriname
Lot, serial or model codes
Lot Number: REQI0632 REQJ0353 REQK0281 REQK0377 RERA0355 RERB0142 RERC0031 RERD0071 RERD0117 RERE0156 RERE0547 RERG0317 RERH0719 RERI0584 RERK0426 RERL0187 RESA0699 RESB0560 RESC0584 RESD0592 RESE0373 RESF0560 RESG0181 RESH0718 RESI0438 RESJ0549 RESJ0550 RETB0903 and RETB0904
510(k) numbers
["K934370"]
PMA numbers
not on file
Product code
FCG
Firm FEI
2020394
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register