Skip to the content

FDA Medical Device Recalls (openFDA)

Ortho-Clinical Diagnostics: VITROS Chemistry Products PHYT Slides, REF/CATALOG 829 8671, Rx ONLY, IVD; Ortho-Clinical Diagnostics, Inc. For in vitro diagnostic use only. VITROS Chemistry Products PHYT Slides quantitatively measure phenytoin (PHYT) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Recall number
Z-0273-2018
Recalling firm
Ortho-Clinical Diagnostics
Product
VITROS Chemistry Products PHYT Slides, REF/CATALOG 829 8671, Rx ONLY, IVD; Ortho-Clinical Diagnostics, Inc. For in vitro diagnostic use only. VITROS Chemistry Products PHYT Slides quantitatively measure phenytoin (PHYT) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Status
Terminated
Initiated
2017-06-12
Posted
not on file
Terminated
2020-08-11
Reason
The firm identified the potential for biased results to be generated when processing VITROS PHYT Slide product with citrate plasma specimen samples. The current Instructions for Use (IFU) and Assay Summary Chart state serum, heparin plasma, and citrate plasma as acceptable specimen types for use with the PHYT assay.
Root cause
Device Design
Action
Ortho-Clinical Diagnostic sent an Urgent Product Correction Letter dated June 12, 2017, sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all customers who have a VITROS analyzer that can process PHYT Slides to inform them that negatively biased results may occur when processing citrate plasma samples using VITROS Chemistry Products PHYT Slides. Customers were instructed to immediately discontinue using citrate plasma samples to process VITROS PHYT Slides. They were also informed that it is still acceptable to use serum and heparin plasma samples to process VITROS PHYT Slides. Foreign affiliates were informed by e-mail on June 12, 2017 of the issue and instructed to notify their consignees of the issue and the appropriate actions to take. For further questions, please call (908) 218-8776.
Quantity
US; 4680; Foreign: 1286
Distribution
Worldwide Distribution - US Distribution including Puerto Rico and the following countries: Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Poland, Singapore, Spain, Russia and Venezuela.
Lot, serial or model codes
Unique Identifier Number ( UDI) 10758750004690; This issue affects expired, current (in date) and future product GENs.
510(k) numbers
["K081543","K941142"]
PMA numbers
not on file
Product code
DIP
Firm FEI
1000305840
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register