Skip to the content

FDA Medical Device Recalls (openFDA)

Philips North America Llc: DigitalDiagnost 4 Chest/Emergency, used to acquire, process, store, display and export digital radiographic images.

Recall number
Z-0272-2024
Recalling firm
Philips North America Llc
Product
DigitalDiagnost 4 Chest/Emergency, used to acquire, process, store, display and export digital radiographic images.
Status
Open, Classified
Initiated
2023-09-21
Posted
2023-11-08
Terminated
not on file
Reason
Potential for units suspended on the ceiling to fail and fall.
Root cause
Under Investigation by firm
Action
The firm distributed recall notifications dated 8/31/2023 via FedEx to affected consignees. Consignees are instructed to immediately stop use of affected devices and contact their local Philips service representative if the Ceiling Suspension becomes unstable, if moving the Ceiling Suspended X-Ray Tube requires a high amount of force, or if moving the Ceiling Suspended X-Ray Tube is noticeably hampered. The recall notice is to be circulated among all device users, so they are aware of this issue and placed as an addendum in the IFU. Consignees are to complete the provided response form and return it to Philips no later than 30 days from notification receipt via email to pd.cnr@philips.com. Philips will contact consignees to schedule a time for a Field Service Engineer to visit customer sites to inspect affected systems and correct the issue. Consignees with any questions are to contact the Customer Care Solutions Center at 1-800-722-9377.
Quantity
13 units
Distribution
Nationwide distribution.
Lot, serial or model codes
Model No. 712029 & 712033; UDI: (01)00884838074521(21) SN19000272, (01)00884838074521(21) SN19000213 , (01)00884838074521(21) 10001077, (01)00884838074521(21) SN20000097, (01)00884838074521(21) SN16000074, (01)00884838074521(21) SN15000238, (01)00884838074521(21) SN16000075, (01)00884838074521(21) SN15000367, (01)00884838074521(21) SN18000428, (01)00884838074521(21) SN17000318, (01)00884838074521(21) SN18000309, (01)00884838074521(21) SN18000204; Serial No. SN15000409, SN19000272, SN19000213, 10001077, SN20000097, SN16000074, SN15000238, SN16000075, SN15000367, SN18000428, SN17000318, SN18000309, SN18000204.
510(k) numbers
["K210692"]
PMA numbers
not on file
Product code
KPR
Firm FEI
3016618143
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register