Skip to the content

FDA Medical Device Recalls (openFDA)

AGFA Corp.: IMPAX CardioVascular (CV) Admin Tool The IMPAX Cardiovascular suite is a cardiovascular information system providing image archiving and image display and modality study-specific structured reporting.

Recall number
Z-0272-2012
Recalling firm
AGFA Corp.
Product
IMPAX CardioVascular (CV) Admin Tool The IMPAX Cardiovascular suite is a cardiovascular information system providing image archiving and image display and modality study-specific structured reporting.
Status
Terminated
Initiated
2011-06-06
Posted
2011-11-23
Terminated
2012-06-26
Reason
Potential for misattribution of data and for a report to become unavailable when the user requests for it to be moved to another study that already has a report associated with it.
Root cause
Software design
Action
An 'URGENT FIELD SAFETY NOTICE" was sent via FED-EX by June 6, 2011 to all the sites using the affected product. The letter described the issue and mitigation. The letter recommends that users not move reports to studies where a report already exists. If a report is being moved because the existing report is irrelevant to the study, then the existing report should be deleted prior to moving the new report to that study. If the report is being moved because the user wants to combine the content of the reports, then the "merge" function should be utilized. The letter also states that a software enhancement is currently available. Acknowledgment via FAX-Back was requested from the sites. Questions regarding the issue should be directed to customers' local Agfa HealthCare Service at 877-777-2432. The firm has identified 185 additional sites and an "URGENT FIELD SAFETY NOTICE" was sent on October 12, 2012.
Quantity
323 units
Distribution
Worldwide Distribution -- USA, Virgin Islands and Canada
Lot, serial or model codes
Software Versions 1.00.26, 1.00.27, CV7.4.SU3 and CV7.8.
510(k) numbers
["K050228"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3001236302
Firm city
Greenville
Firm state
SC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register