AGFA Corp.: IMPAX CardioVascular (CV) Admin Tool The IMPAX Cardiovascular suite is a cardiovascular information system providing image archiving and image display and modality study-specific structured reporting.
- Recall number
- Z-0272-2012
- Recalling firm
- AGFA Corp.
- Product
- IMPAX CardioVascular (CV) Admin Tool The IMPAX Cardiovascular suite is a cardiovascular information system providing image archiving and image display and modality study-specific structured reporting.
- Status
- Terminated
- Initiated
- 2011-06-06
- Posted
- 2011-11-23
- Terminated
- 2012-06-26
- Reason
- Potential for misattribution of data and for a report to become unavailable when the user requests for it to be moved to another study that already has a report associated with it.
- Root cause
- Software design
- Action
- An 'URGENT FIELD SAFETY NOTICE" was sent via FED-EX by June 6, 2011 to all the sites using the affected product. The letter described the issue and mitigation. The letter recommends that users not move reports to studies where a report already exists. If a report is being moved because the existing report is irrelevant to the study, then the existing report should be deleted prior to moving the new report to that study. If the report is being moved because the user wants to combine the content of the reports, then the "merge" function should be utilized. The letter also states that a software enhancement is currently available. Acknowledgment via FAX-Back was requested from the sites. Questions regarding the issue should be directed to customers' local Agfa HealthCare Service at 877-777-2432. The firm has identified 185 additional sites and an "URGENT FIELD SAFETY NOTICE" was sent on October 12, 2012.
- Quantity
- 323 units
- Distribution
- Worldwide Distribution -- USA, Virgin Islands and Canada
- Lot, serial or model codes
- Software Versions 1.00.26, 1.00.27, CV7.4.SU3 and CV7.8.
- 510(k) numbers
- ["K050228"]
- PMA numbers
- not on file
- Product code
- LLZ
- Firm FEI
- 3001236302
- Firm city
- Greenville
- Firm state
- SC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28