Skip to the content

FDA Medical Device Recalls (openFDA)

Richard Wolf Medical Instruments Corp.: Richard Wolf Hulka Clip; for single patient use only; a sterile tubal occluding device; 2 clips per procedure tray/pouch, 10 pouches per shipping carton; Richard Wolf Medical Instruments Corp., Vernon Hills, IL 60061; item 4986.90. Indicated for use for female sterilization (permanent contraception) by occluding the fallopian tubes

Recall number
Z-0272-2010
Recalling firm
Richard Wolf Medical Instruments Corp.
Product
Richard Wolf Hulka Clip; for single patient use only; a sterile tubal occluding device; 2 clips per procedure tray/pouch, 10 pouches per shipping carton; Richard Wolf Medical Instruments Corp., Vernon Hills, IL 60061; item 4986.90. Indicated for use for female sterilization (permanent contraception) by occluding the fallopian tubes
Status
Terminated
Initiated
2009-10-16
Posted
2009-11-17
Terminated
2011-03-16
Reason
The sterile pouch seal may fail to remain intact during handling rendering the Hulka clip non-sterile.
Root cause
Packaging
Action
Richard Wolfe sent recall letters dated October 16, 2009 to the direct accounts via certified mail on the same date. The accounts were informed that the lot of Hulka clips did not meet their sterile seal burst test requirements, and that the sterile seal may fail to remain intact during handling, resulting in the clips becoming contaminated or non-sterile. They were instructed to inspect their inventory for the affected lot, quarantine any product found and call Richard Wolf at 800-323-9653, ext 225 or 380 to arrange for the return of the affected product for replacement. The accounts were also requested to complete and return the enclosed response form to Richard Wolf, even if they do not have any of the affected lot in stock.
Quantity
150 cartons
Distribution
Nationwide
Lot, serial or model codes
lot number 1391, expiry date 02.2010
510(k) numbers
not on file
PMA numbers
["P870080"]
Product code
HGB
Firm FEI
1418479
Firm city
Vernon Hills
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register