FDA Medical Device Recalls (openFDA)
Primus Medical LLC: AC Powered adjustable hospital bed, Prime Bed TL500, Distributed by: Primus Medical LLC, 8401 Southern Blvd, Boardman, OH 44512. Usage: Hospital Bed.
- Recall number
- Z-0270-2013
- Recalling firm
- Primus Medical LLC
- Product
- AC Powered adjustable hospital bed, Prime Bed TL500, Distributed by: Primus Medical LLC, 8401 Southern Blvd, Boardman, OH 44512. Usage: Hospital Bed.
- Status
- Terminated
- Initiated
- 2011-02-02
- Posted
- 2012-11-08
- Terminated
- 2014-05-14
- Reason
- The "Retaining washers & Clevis Pins" provided with the AC -powered adjustable hospital beds may fail under certain circumstances. The pins at the top of the leg were disengaging from the channel. This may cause the bed frame to come off tracks or collapse.
- Root cause
- Equipment maintenance
- Action
- The firm, Primus Medical LLC, sent an "Important Notification Relating to Primus Medical Product TL500" letter dated March 29, 2011 to their customers. The letter described the product, problem and actions to be taken. The customers were instructed to replace the "Retaining washers & Clevis Pins" with the "Clevis Pins & Bow Tie Cotter Pins" in the TL500-001 Service Pack provided with the letter in order to avoid any potential failure and/or bed collapse and to return the" Retaining washers & Clevis Pins" after the replacement procedure is performed. If you have any questions, or need assistance in performing the replacement, call 330-965-9681.
- Quantity
- 342 Beds
- Distribution
- Nationwide distribution: USA including states of: AZ, IL, KY, MD, MI, OH, OR, PA, TX and VA.
- Lot, serial or model codes
- Model #TL500; Serial Number: 001, 002, 003, 004, 005, 006, 007, 010, 013, 014, 015, 016, 017, 018, 021, 022, 023, 024, 025, 026, 027, 028, 029, 030, 031, 032, 033, 034, 035, 039, 040, 041, 042, 043, 044, 045, 046, 047, 048, 049, 050, 052, 053, 054, 055, 056, 057, 058, 059, 060, 061, 062, 063, 064, 065, 070, 075, 081, 09350014, 09350015, 09350016, 09350017, 09350018, 09350021, 09350023, 09350025, 09350026, 09350027, 09350028, 09350029, 09350030, 09350031, 09350032, 09350033, 09350035, 09350038, 09350040, 09350041, 09350042, 09350045, 09350048, 09350049, 09350054, 09350055, 09350056, 09350057, 09350058, 09350059, 09350060, 09350062, 09350063, 09350064, 09350065, 09350066, 09350067, 09350068, 09350069, 09350070, 09350072, 09350074, 09350082, 09350085, 09350087, 09350088, 09350089, 09350090, 09350091, 09350092, 09430001, 09430002, 09430003, 09430004, 09430005, 09430006, 09430007, 09430008, 09430010, 09430013, 09430014, 09430015, 09430017, 09430019, 09430020, 09430021, 09430022, 09430024, 09430026, 09430027, 09430028, 09430029, 09430030, 09430032, 09430033, 09430034, 09430036, 09430037, 09430038, 09430039, 09430040, 09430041, 09430042, 09430043, 09430044, 09430045, 09430046, 09430047, 09430049, 09430050, 09430051, 09430052, 09430053, 09430054, 09430056, 09430057, 09430058, 09430059, 09430060, 09430061, 09430062, 09430063, 09430065, 09430066, 09463067, 09430068, 09430069, 09430070, 09430071, 09430072, 09430074, 09430075, 09430076, 09430077, 09430078, 09430079, 09430080, 09430081, 09430083, 09430085, 09430086, 09430087, 09430088, 09430089, 09430091, 09430093, 09430094, 09430097, 09430099, 09430100, 09520001, 09520002, 09520003, 09520004, 09520005, 09520006, 09520007, 09520008, 09520009, 09520011, 09520012, 09520013, 09520015, 09520017, 09520018, 09520019, 09520020, 09520021, 09520022, 09520025, 09520026, 09520027, 09520028, 09520029, 09520030, 09520031, 09520032, 09520033, 09520034, 09520035, 09520036, 09520037, 09520038, 09520039, 09520040, 09520041, 09520042, 09520043, 09520044, 09520045, 09520046, 09520047, 09520048, 09520049, 09520050, 10040001, 10040003, 10040004, 10040005, 10040006, 10040007, 10040008, 10040009, 10040010, 10040011, 10040012, 10040013, 10040014, 10040015, 10040016, 10040017, 10040028, 10040029, 10040030, 10040032, 10040033, 10040034, 10040035, 10040036, 10040037, 10040038, 10040039, 10040040, 10040041, 10040042, 10040043, 10040044, 10040045, 10040046, 10040047, 10040048, 10040049, 10040050, 10040051, 10040052, 10040053, 10040054, 10040055, 10040056, 10040057, 10040058, 10040059, 10040062, 10040063, 10040064, 10040065, 10040066, 10040067, 10040068, 10040069, 10040070, 10040071, 10040073, 10040074, 10040075, 10040076, 10040077, 10040078, 10400029, 10400050, 10400056, 10400057, 10400058, 10400059, 10400065, 10370016, 10370017, 10370018, 10370019, 10370020, 10370032, 10370033, 10370034, 10370035, 10370038, 10370052, 10370053, 10370054, 10370056, 10370060, 10370063, 10370078, 10370079, 10370080 , 09380031, 09380029, 09380064, 09380070, 09380089, 09380040, 09380038, 09380085, 09380045, 09380054, 09380060, 09380059, 09380058, 09380048, 09380042, 09380038, 09380069, 09380063, 09380067 & 09380065.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FNL
- Firm FEI
- 3007538326
- Firm city
- Boardman
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28