Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Biomet, Inc.: ExploR Radial Head Replacement System. PN: 11-210061 LN: 243610. The stem is placed inside a polyurethane bag. The bagged component will then be placed in an inner PETG tray. The screw will be placed inside a separate polybag, and placed in the PETG tray with foam pieces added to the top and bottom of the tray for protection. The inner tray and the contents within will then have a Tyvek lid heat sealed to the tray. The sealed inner tray will be fit into an outer PETG tray and a Tyvek lid heat sealed to the tray. The entire sealed unit is placed in a SBS carton that will be shrink wrapped for protection.

Recall number
Z-0269-2018
Recalling firm
Zimmer Biomet, Inc.
Product
ExploR Radial Head Replacement System. PN: 11-210061 LN: 243610. The stem is placed inside a polyurethane bag. The bagged component will then be placed in an inner PETG tray. The screw will be placed inside a separate polybag, and placed in the PETG tray with foam pieces added to the top and bottom of the tray for protection. The inner tray and the contents within will then have a Tyvek lid heat sealed to the tray. The sealed inner tray will be fit into an outer PETG tray and a Tyvek lid heat sealed to the tray. The entire sealed unit is placed in a SBS carton that will be shrink wrapped for protection.
Status
Terminated
Initiated
2017-06-16
Posted
not on file
Terminated
2018-09-06
Reason
Zimmer Biomet is conducting a medical device recall for the ExploR 6x24mm Implantable Stem with Screw due to a complaint alleging that the Modular Radial HD lock screw was not included in the package for the stem.
Root cause
Packaging change control
Action
Zimmer Biomet sent US Fiedl action notices 6/16/17 via email and fed Ex. On 9/5/17, firm expanded their recall to include 1 lot and distributed Amended notices via email and FedEx
Quantity
not on file
Distribution
US, Canada, OUS.
Lot, serial or model codes
Part Lot UDI Number 11-210061 243610 and 202900; (01)00880304210448(17)260817(10)243610
510(k) numbers
["K051385"]
PMA numbers
not on file
Product code
KWI
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register