Skip to the content

FDA Medical Device Recalls (openFDA)

Guidant Corporation: Guidant CONTAK RENEWAL 3 RF (models H210, H215), CONTAK RENEWAL 3 RF HE (models H217, H219) ,CONTAK RENEWAL 4 RF (model H230), and CONTAK RENEWAL 4 RF HE (model H239) cardiac resynchronization therapy defibrillator (CRT-D). Guidant Corporation, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA.

Recall number
Z-0269-2007
Recalling firm
Guidant Corporation
Product
Guidant CONTAK RENEWAL 3 RF (models H210, H215), CONTAK RENEWAL 3 RF HE (models H217, H219) ,CONTAK RENEWAL 4 RF (model H230), and CONTAK RENEWAL 4 RF HE (model H239) cardiac resynchronization therapy defibrillator (CRT-D). Guidant Corporation, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA.
Status
Terminated
Initiated
2006-09-30
Posted
2006-12-16
Terminated
2007-04-01
Reason
Shock Effectiveness: there is a potential for malfunction of a high voltage wire, which could compromise effectiveness of shock therapy. While no energy is lost, defibrillation thresholds may be higher (and safety margin may be reduced) because the shocking vector has changed.
Root cause
Other
Action
An Important Medical Device Information letter, dated 10/09/2006 was sent to physicians following patients with a potentially affected device. The letter describes the issue, provides for a projected rate of occurrence, and provides recommendations. An October 9, 2006 Product Advisory Letter, beginning October 30, 2006, was also sent to provide an update that further laboratory testing and analysis indicates that the nominal probability of malfunction is higher than their earlier prediction. Non-implanted devices have been retrieved.
Quantity
57 (46-USA, 11 Foreign- Renewal 4RF)
Distribution
Worldwide distribution ---- including states of CA, FL, GA, IA, IL, KY, MA, MO, NJ, NY, OH, PA, TX and countries of Italy, Netherlands, Sweden, United Kingdom.
Lot, serial or model codes
Model H210, serial numbers: 204765, 204766, 204767, 204769, 204770, 204771, 204772, 204773, 204774, 204777, 204784, 205492, 205494, 205495, 205497, 205498, 205504, 205507, 205512, 205523, 205525.   Model H215, serial numbers: 703984, 703985, 703986, 704415, 704417, 704418, 704421, 704422, 704423, 704424, 704425.   Model H217, serial numbers: 704509, 704531, 704534, 704545, 704547, 704550, 705267, 705294, 705335, 705350.   Model H219, serial numbers: 203320, 203344, 203355, 203377, 203382.   Model H230, serial numbers: 201149, 201150, 201151, 201152, 201153, 201154, 201155.   Model H239, serial numbers: 650451, 650453, 650454.
510(k) numbers
not on file
PMA numbers
["P010012-Rene"]
Product code
LWP
Firm FEI
2124215
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register