Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Perfusion Systems: Medtronic Perfusion Tubing Packs, Models TL7M64R3, TL7P17R1 & HY2W85R9, Sterilized using Ethylene Oxide. The Tubing Pack design and components are specified by the user, they are packaged with consideration of setup efficiency and respect for the environment, and they are intended for use in the extracorporeal circuit during cardiopulmonary bypass procedures. Medtronic Tubing Packs can contain components with Carmeda BioActive Surface, Trillium BioSurface, Balance BioSurface or no coating. Sterile barrier for these packs is provided either by a Tyvek lid seal on the tray or by a header bag.

Recall number
Z-0268-2016
Recalling firm
Medtronic Perfusion Systems
Product
Medtronic Perfusion Tubing Packs, Models TL7M64R3, TL7P17R1 & HY2W85R9, Sterilized using Ethylene Oxide. The Tubing Pack design and components are specified by the user, they are packaged with consideration of setup efficiency and respect for the environment, and they are intended for use in the extracorporeal circuit during cardiopulmonary bypass procedures. Medtronic Tubing Packs can contain components with Carmeda BioActive Surface, Trillium BioSurface, Balance BioSurface or no coating. Sterile barrier for these packs is provided either by a Tyvek lid seal on the tray or by a header bag.
Status
Terminated
Initiated
2015-10-28
Posted
not on file
Terminated
2016-08-17
Reason
This recall is being initiated due to a potential breach of the Tubing Pack Header Bags manufactured in these select lots. The seal in question is between the Tyvek and the Poly side of the header bag.
Root cause
Packaging process control
Action
Medtronic sent an Urgent Medical Device Recall letter dated October 2015 to all affected consignees. The letter described the problem and the product involved in the recall. Consignees were advised to quarantine and return all affected product and to complete and return the Customer Confirmation certificate. For questions contact your Medtronic representative or contact Medtronic's national Answering service at 800-633-8766.
Quantity
496
Distribution
Nationwide Distribution including US: AL, IL, IN, and WI.
Lot, serial or model codes
Model Lot Numbers Lot Number HY2W85R9 209704344,209704345,209704346,  209704347,210010558    TL7M64R3 209789393 TL7P17R1 209789333
510(k) numbers
["K012538"]
PMA numbers
not on file
Product code
DWF
Firm FEI
1000116158
Firm city
Brooklyn Park
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register