FDA Medical Device Recalls (openFDA)

Kerr Corporation: The brand name of the device is AlgiNot, a dental impression material. Product Part Numbers AlgiNot Intra Kit 34683 AlgiNot Volume Refills 34682 Lot Number 1-1017 AlgiNot is a Class II Medical Device. The device listing number this product is 0136496 and the establishment registration number for Kerr Corporation is 1815757. Intended as an alternative to traditional alginate materials.

Recall number
Z-0268-2013
Recalling firm
Kerr Corporation
Product
The brand name of the device is AlgiNot, a dental impression material. Product Part Numbers AlgiNot Intra Kit 34683 AlgiNot Volume Refills 34682 Lot Number 1-1017 AlgiNot is a Class II Medical Device. The device listing number this product is 0136496 and the establishment registration number for Kerr Corporation is 1815757. Intended as an alternative to traditional alginate materials.
Status
Terminated
Initiated
2012-02-21
Posted
2012-11-08
Terminated
2012-11-15
Reason
The firm recalled the product because it may cure harder than expected within the shelf life, which could result in a material that may be more difficult to remove. Use of this product on patients with loose teeth may result in possible tooth fracture or tooth extraction.
Root cause
Nonconforming Material/Component
Action
Kerr Corporation sent "URGENT: MEDICAL DEVICE RECALL" letters to all affected customers. The letters identified the product, problem, and actions to be taken by the customers. US, Canadian, and Rest of World consignees were notified via USPS 1st class mail on June 19, 2012, and Australian consignees on July 5, 2012. Consignees were instructed to complete and return the Acknowledgement Form. Notification of the recall expansion was sent, via USPS 1st class mail, on February 21, 2012 to US, Canadian, Australia and European, consignees; however, once further investigation determined that all unexpired lots of the product were affected, the Alginot recall was expanded a second recall letter was sent to clarify that all lots of AlgiNot FS Volume recall and AlgiNot recall expansion are affected by this expanded recall. Letters were mailed to US, Canadian and Rest of World (ROW) customers on June 19, 2012; Australian and European consignees were notified of the expansion on July 5, 2012. All consignees were instructed to complete the Return Form to acknowledge the updated instructions for use and the added cautionary statement. Consignees not responding are to be contacted at least three times via telephone/fax/e-mail as follow up on the recall.
Quantity
45 kits and 87 refills
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada, Europe, and Australia.
Lot, serial or model codes
¿Product Part Numbers Lot Number  AlgiNot Intra Kit 34683 1-1017  AlgiNot Volume Refills 34682 1-1017
510(k) numbers
["K050604"]
PMA numbers
not on file
Product code
ELW
Firm FEI
1815757
Firm city
Romulus
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register