Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries Inc: Medline Premium Wet Skin Prep Tray, Sterile, Latex Free; a single use surgical convenience tray, EO sterilized; Reorder DYND70660; 20 trays per case; Medline Industries, Inc., Mundelein, IL 60060-4486

Recall number
Z-0268-2007
Recalling firm
Medline Industries Inc
Product
Medline Premium Wet Skin Prep Tray, Sterile, Latex Free; a single use surgical convenience tray, EO sterilized; Reorder DYND70660; 20 trays per case; Medline Industries, Inc., Mundelein, IL 60060-4486
Status
Terminated
Initiated
2006-11-21
Posted
2006-12-14
Terminated
2008-01-31
Reason
The product, labeled as sterile, had not gone through the sterilization process at the time of shipment.
Root cause
Other
Action
On 11/21/06, Medline telephoned the accounts who may have received the recalled lots, informing them that the skin scrub trays may not have gone through the full manufacturing cycle and are not sterile. They were requested to check their inventories for the two affected lots, removing any found from inventory and placing it on hold for return to Medline. The accounts were then sent a recall letter via fax on the same date, reiterating the telephone calls, and requesting the accounts to complete and fax back the attached response form, indicating amount of each lot placed on hold. Any questions were directed to Medline''s recall hotline at 866-359-1704.
Quantity
47 cases
Distribution
Minnesota, Illinois, Wisconsin and Indiana
Lot, serial or model codes
Reorder # DYND70660, lots 06JC0016 and 06JC0017, exp. 08-08
510(k) numbers
["K914174"]
PMA numbers
not on file
Product code
LRO
Firm FEI
1417592
Firm city
Mundelein
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register