Skip to the content

FDA Medical Device Recalls (openFDA)

C.R. Bard, Inc.: I-125 CE Marked QuickLink Cartridges; Sterile; Bard Brachysource Radioactive Seeds in QuickLink Cartridges Indicated for permanent interstitial treatment of selected localized tumors such as: head and neck, lung, pancreas, and early stage prostate. Implants are indicated to treat residual tumors following completion of a course of external radiation therapy and for recurrent tumors.

Recall number
Z-0267-2018
Recalling firm
C.R. Bard, Inc.
Product
I-125 CE Marked QuickLink Cartridges; Sterile; Bard Brachysource Radioactive Seeds in QuickLink Cartridges Indicated for permanent interstitial treatment of selected localized tumors such as: head and neck, lung, pancreas, and early stage prostate. Implants are indicated to treat residual tumors following completion of a course of external radiation therapy and for recurrent tumors.
Status
Terminated
Initiated
2016-12-09
Posted
not on file
Terminated
2020-04-24
Reason
Activity and quantity of seeds sent in shipments were incorrect.
Root cause
Other
Action
Customers were notified via letter on December 9, 2016. In the case of both customers, the hospital utilized the product to complete the scheduled implants. Adjustments were made to the dose plans and the number of seeds implanted. According to C.R. Bard, the physicians would evaluate the risk by taking into account each individual patient's condition, comorbidities, and treatment plans, and would adjust the dose. For further questions, please call (770) 784-6120.
Quantity
134 seeds
Distribution
Belgium & South Africa.
Lot, serial or model codes
Product Code: 1251QCSCE, Lot No. BBAY0072, BBAY0079, BBAY0090, BBAY0071, BBAY0073.
510(k) numbers
["K043246"]
PMA numbers
not on file
Product code
KXK
Firm FEI
1018233
Firm city
Covington
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register