Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diagnostics Operations, Inc.: Cobas u 601 urinalysis test system, Catalog number 06334601001 The cobas U 601 urinalysis test system is comprised of the cobas u 601 urine analyzer and the cobas u pack. The cobas u 601 urine analyzer when used with the cobas u pack is a fully automated urinalysis system intended for the in vitro qualitative or semi-quantitative determination of urine analytes, including pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, color and erythrocytes, as well as clarity.

Recall number
Z-0266-2022
Recalling firm
Roche Diagnostics Operations, Inc.
Product
Cobas u 601 urinalysis test system, Catalog number 06334601001 The cobas U 601 urinalysis test system is comprised of the cobas u 601 urine analyzer and the cobas u pack. The cobas u 601 urine analyzer when used with the cobas u pack is a fully automated urinalysis system intended for the in vitro qualitative or semi-quantitative determination of urine analytes, including pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, color and erythrocytes, as well as clarity.
Status
Terminated
Initiated
2021-10-18
Posted
not on file
Terminated
2023-05-31
Reason
A potential risk for false negative nitrite results exists when endogenous creatinine levels are 15,000 mg/L and above. No interference of Nitrite results was observed at creatinine levels up to 9,000 mg/L.
Root cause
Software design
Action
On October 18, 2021, the firm, Roche, sent "URGENT MEDICAL DEVICE CORRECTION" letters to affected consignees. Customers were notified of the potential for false negative nitrite test results with elevated endogenous creatinine concentrations. Customers were provided with an amended cobas u pack method sheet that provides the updated interference claim for endogenous creatinine. Nitrite testing is commonly included in the process of diagnosing urinary tract infections. In case of false negative nitrite results, particularly in asymptomatic patients or patient with nonspecific symptoms, further diagnostic measures in urine might be skipped, resulting in an increased potential risk for disease progression and complications of UTI. Actions required: - Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. - Review the updated cobas u pack method sheet included with the letter and update your laboratory's procedures, if needed. - Complete and return the customer response form via email to roche6051@sedgwick.com. - File the recall notice for future reference. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
Quantity
636
Distribution
US Nationwide distribution.
Lot, serial or model codes
UDI 04015630932252
510(k) numbers
["K183432"]
PMA numbers
not on file
Product code
KQO
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register