Skip to the content

FDA Medical Device Recalls (openFDA)

Medacta Usa Inc: The GMK knee system: designed for cemented use in total knee arthroplasty. Sphere and Primary Inserts with part number ranges: 02.12.0110FL - 02.12.0620FL; 02.12.0110FR - 02.12.0620FR; 02.07.0110PSF - 02.07.0620PSF

Recall number
Z-0266-2018
Recalling firm
Medacta Usa Inc
Product
The GMK knee system: designed for cemented use in total knee arthroplasty. Sphere and Primary Inserts with part number ranges: 02.12.0110FL - 02.12.0620FL; 02.12.0110FR - 02.12.0620FR; 02.07.0110PSF - 02.07.0620PSF
Status
Terminated
Initiated
2017-10-24
Posted
2017-12-20
Terminated
2018-06-20
Reason
Medacta has observed a worldwide occurrence rate of post-operative insert screw back-out of approximately 0.1%. For this reason, we stress the importance of using the 3.5 Nm torque limiting driver if the screw is used, as directed in our surgical techniques.
Root cause
Device Design
Action
All surgeon customers shall be notified via email by October 24, 2017 of the reminder to follow the surgical technique. To ensure that surgeons receive the notification, we will send the letter via email so that we can track and ensure the message was received. If the email is returned, we would then send a personalized letter directly to the surgeon. A copy of this notification is included with the submission of this report.
Quantity
12,768 units
Distribution
nationwide
Lot, serial or model codes
all serial numbers
510(k) numbers
["K090988"]
PMA numbers
not on file
Product code
JWH
Firm FEI
3006639916
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register