Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc: Access Immunoassay Systems, 862.2160 discrete photometric chemistry analyzer.

Recall number
Z-0265-05
Recalling firm
Beckman Coulter Inc
Product
Access Immunoassay Systems, 862.2160 discrete photometric chemistry analyzer.
Status
Terminated
Initiated
2004-11-08
Posted
2004-11-30
Terminated
2012-08-13
Reason
A rare condition of the Access 2 Immunoassay Systems software versions 2.0, 2.1 and 2.2 could present a risk of an erroneous confirmatory result under a specific set of circumstances. The Chlamydia Blocking results may be affected. The instrument will erroneously perform the blocking calculation on the affected test by using the mean of the diluted and neat results.
Root cause
Other
Action
Product Corrective Action (PCA) letter that informs customers (who ordered Access Chlamydia reagent in the past 24 months and have an Access 2 Immunoassay System) of this potential issue sent via US mail on November 8, 2004. The letter instructs customers to use unique Sample IDs for all samples assayed on the system - including those which have been manipulated offline prior to assay. A Technical Information Update will be issued to inform all support personnel of this issue.
Quantity
1,197
Distribution
Nationwide and Canada
Lot, serial or model codes
Access 2 Immunoassay Systems Part Number 81600N software versions 20., 2.1, and 2.2. Access Immunoassay Systems Chlamydia Reagent Kit Part Number 34401.
510(k) numbers
["K922823"]
PMA numbers
not on file
Product code
CGN
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register