FDA Medical Device Recalls (openFDA)
POM Medical LLC: POM Procedural Oxygen Masks, Endoscopic, Conscious Sedation, Bronchoscopy Mask, as follows: (1) POM Medical, LLC, REF 1001-MM, High/Med Concentrations, Distributed by POM Medical, LLC, Moorpark, CA; (2) curaplex, REF 301-0318LT, Medium Concentration, Distributed by Sarnova HC, LLC's family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency Medical Products, Inc. & Tri-anim Health Services, Inc., Dublin, OH; and (3) POM Medical, LLC REF 1001-MF, High/Med Concentrations, Distributed by POM Medical, LLC, Moorpark, CA. All are packaged 1 mask/plastic pouch, 30 masks/case.
- Recall number
- Z-0262-2020
- Recalling firm
- POM Medical LLC
- Product
- POM Procedural Oxygen Masks, Endoscopic, Conscious Sedation, Bronchoscopy Mask, as follows: (1) POM Medical, LLC, REF 1001-MM, High/Med Concentrations, Distributed by POM Medical, LLC, Moorpark, CA; (2) curaplex, REF 301-0318LT, Medium Concentration, Distributed by Sarnova HC, LLC's family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency Medical Products, Inc. & Tri-anim Health Services, Inc., Dublin, OH; and (3) POM Medical, LLC REF 1001-MF, High/Med Concentrations, Distributed by POM Medical, LLC, Moorpark, CA. All are packaged 1 mask/plastic pouch, 30 masks/case.
- Status
- Terminated
- Initiated
- 2018-07-11
- Posted
- not on file
- Terminated
- 2021-01-04
- Reason
- Complaints were received on the product, such as the re-breather bag would not fill/inflate or the oxygen line pops off after the oxygen is turned on due to a clogged port.
- Root cause
- Nonconforming Material/Component
- Action
- The firm initiated recall by telephone beginning 7/11/2018 requesting return of the product.
- Quantity
- 5,610 units
- Distribution
- Distribution was made to CA, FL, MO, OH, and PA.
- Lot, serial or model codes
- Model 1001-MM - Lot numbers 021418-021421, 032218-032221, 032618-032621, and 032818-032821; Model KE1001-MM, an internal number used by one of the distributors (same as 1001-MM) - Lot numbers 021418-021421 and 032818-032821; Model 301-0318LT - Lot numbers 022218-022221, 032218-032221, 032618-032621; and Model 1001-MF - Lot number 032818-032821.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- BYG
- Firm FEI
- 3010411911
- Firm city
- Moorpark
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as