Skip to the content

FDA Medical Device Recalls (openFDA)

POM Medical LLC: POM Procedural Oxygen Masks, Endoscopic, Conscious Sedation, Bronchoscopy Mask, as follows: (1) POM Medical, LLC, REF 1001-MM, High/Med Concentrations, Distributed by POM Medical, LLC, Moorpark, CA; (2) curaplex, REF 301-0318LT, Medium Concentration, Distributed by Sarnova HC, LLC's family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency Medical Products, Inc. & Tri-anim Health Services, Inc., Dublin, OH; and (3) POM Medical, LLC REF 1001-MF, High/Med Concentrations, Distributed by POM Medical, LLC, Moorpark, CA. All are packaged 1 mask/plastic pouch, 30 masks/case.

Recall number
Z-0262-2020
Recalling firm
POM Medical LLC
Product
POM Procedural Oxygen Masks, Endoscopic, Conscious Sedation, Bronchoscopy Mask, as follows: (1) POM Medical, LLC, REF 1001-MM, High/Med Concentrations, Distributed by POM Medical, LLC, Moorpark, CA; (2) curaplex, REF 301-0318LT, Medium Concentration, Distributed by Sarnova HC, LLC's family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency Medical Products, Inc. & Tri-anim Health Services, Inc., Dublin, OH; and (3) POM Medical, LLC REF 1001-MF, High/Med Concentrations, Distributed by POM Medical, LLC, Moorpark, CA. All are packaged 1 mask/plastic pouch, 30 masks/case.
Status
Terminated
Initiated
2018-07-11
Posted
not on file
Terminated
2021-01-04
Reason
Complaints were received on the product, such as the re-breather bag would not fill/inflate or the oxygen line pops off after the oxygen is turned on due to a clogged port.
Root cause
Nonconforming Material/Component
Action
The firm initiated recall by telephone beginning 7/11/2018 requesting return of the product.
Quantity
5,610 units
Distribution
Distribution was made to CA, FL, MO, OH, and PA.
Lot, serial or model codes
Model 1001-MM - Lot numbers 021418-021421, 032218-032221, 032618-032621, and 032818-032821;  Model KE1001-MM, an internal number used by one of the distributors (same as 1001-MM) - Lot numbers 021418-021421 and 032818-032821;  Model 301-0318LT - Lot numbers 022218-022221, 032218-032221, 032618-032621; and  Model 1001-MF - Lot number 032818-032821.
510(k) numbers
not on file
PMA numbers
not on file
Product code
BYG
Firm FEI
3010411911
Firm city
Moorpark
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register