NXTHERA: Rezum System, Vapor BPH Ablation System, Model Number D2201-003. NxThera Inc. 7351 Kirkwood Lane Suite 138 Maple Grove, MN 55369 USA. Intended to relieve symptoms, obstructions, and reduce prostate tissue associated with benign prostatic hyperplasia (BPH). It is indicated for men greater or equal to 50 years of age with a prostate volume greater than or equal to 30cm3 and less than or equal to 80cm3. The Rezum System is also indicated for treatment of prostates with hyperplasia of the central zone and/or a median lobe.

Recall number
Z-0262-2018
Recalling firm
NXTHERA
Product
Rezum System, Vapor BPH Ablation System, Model Number D2201-003. NxThera Inc. 7351 Kirkwood Lane Suite 138 Maple Grove, MN 55369 USA. Intended to relieve symptoms, obstructions, and reduce prostate tissue associated with benign prostatic hyperplasia (BPH). It is indicated for men greater or equal to 50 years of age with a prostate volume greater than or equal to 30cm3 and less than or equal to 80cm3. The Rezum System is also indicated for treatment of prostates with hyperplasia of the central zone and/or a median lobe.
Status
Terminated
Initiated
2017-07-28
Posted
not on file
Terminated
2018-08-01
Reason
Needle Bond may not be able to withstand the tension force applied when the shaped needle is deployed into and retracted from prostate tissue to deliver treatments.
Root cause
Nonconforming Material/Component
Action
Customers were notified initially by sales representatives on or about July 18, 2017. Any affected devices were to be returned to NxThera. Additionally, a formal recall letter was mailed to customers on approximately July 28, 2017. NxThera's telephone contact information and FEDEX number will be provided on the verification form. For further questions, please call (763) 515-2084..
Quantity
113 units total
Distribution
USA Distribution to the states of : AL, AZ, CA, CO, FL, GA, IL, MA, MI, MN, MO, NE, NJ, NY, OH, PA, TN, UT, WA.
Lot, serial or model codes
UDI 08855757006010. Lot Number of Outside Cases (Pack Lot) 2017020920 with Lot Number of Inside Boxes (Product Lot) 2017011612 and Exp Date 2018-01-31: Lot Number of Outside Cases (Pack Lot) 2017030286 with Lot Number of Inside Boxees (Product Lot) 2017011187 and Exp Date 2018-01-31.
510(k) numbers
["K160417"]
PMA numbers
not on file
Product code
KNS
Firm FEI
3007795803
Firm city
Maple Grove
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register