Implant Direct Sybron Manufacturing, LLC: Implant Direct Scan Adapter Assembly Replant platform for the following part numbers: 6035-09PT-M (3.5mmD)/UDI : 10841307123934,;6043-09PT-M (4.3mmD)/UDI : 10841307123941; and 6050-09PT-M (5.00mmD)/ UDI: 10841307123958
- Recall number
- Z-0261-2020
- Recalling firm
- Implant Direct Sybron Manufacturing, LLC
- Product
- Implant Direct Scan Adapter Assembly Replant platform for the following part numbers: 6035-09PT-M (3.5mmD)/UDI : 10841307123934,;6043-09PT-M (4.3mmD)/UDI : 10841307123941; and 6050-09PT-M (5.00mmD)/ UDI: 10841307123958
- Status
- Terminated
- Initiated
- 2018-09-06
- Posted
- not on file
- Terminated
- 2021-07-01
- Reason
- The action is being initiated due to incorrect assembly of the scan adapter . The result of the incorrect assembly could cause the clinician not to be able to capture the position and orientation of a dental implant(s) or lab analog during the digital scanning process which could result in a delay for the patient.
- Root cause
- Nonconforming Material/Component
- Action
- On X/xx/xxx, consignees were mailed a "Urgent: Medical Device Recall" letter via XXX. The recall notification asked customers to do the following: 1. Please review your inventory for the affected product. 2. Please complete and return the Acknowledgement Form within 48 hours for the product listed above; Quarantine product and return product listed above. 3. If you are an authorized Implant Direct Sybron Manufacturing LLC distributor, we request that you identify those customers that may have been shipped the affected product lot and contact these customers to inform them of this issue within fortyeight (48) hours of receipt of this notification. 4. If you have any of the affected product, please return the product and we will send you a replacement part. If you have any questions contact Implant Direct Sybron Manufacturing LLC Customer Care at 1-888-649-6425. Customer Care is available Monday through Friday, between the hours of 5am-5pm PST. 5.If you or your patients has experienced any issues as a result of the affected products noted in this communication, you may voluntarily report the incident to the FDA through the MEDWATCH reporting system at the following: http://www.fda.gov/Safety/MedWatch/HowToReport/ucm053074.htm
- Quantity
- 107 Scan Adapters
- Distribution
- US: OH, CO, UT, NY, PA, CA, FL , GA, LA, NE, NJ,SC, MA, VA, UT, OUS: Canada, Australia,
- Lot, serial or model codes
- Part numbers/Lot Numbers: 6035-09PT/Lot # 104539; 6043-09PT/ Lot # 104540; 6050-09PT/Lot # 104511
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- NDP
- Firm FEI
- 3001617766
- Firm city
- Westlake Village
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28